background image

8

9. Each quantity is expressed in keV FWHM. The
quantity N

det

 is known as the "noise width" of the

detector, and is included as one of the specified
parameters.  By use of the above equation and with
a knowledge of the noise of the preamplifier, the
noise width of the detector can be determined. The
significance of this noise width in evaluating the
detector is subject to interpretation, but generally
the actual resolution of the detector is related
directly to the noise width in its normal applications.
The most useful purpose for determining this
quantity for the detector is occasional monitoring of
the detector noise width to verify that its
characteristics have not undergone any significant
change during use. 

10. Use an ORTEC 419 Precision Pulse Generator
with a matched charge termination to measure the
rise time of the 142PC through the T  or E output
connector. Connect the 419 output through the
charge terminator to the input of the 142PC and use
an oscilloscope with a fast (1 ns if possible) rise
time. The rise time of the preamplifier can then be
computed by: 

     (Total risetime)

2

 = (Preamp risetime)

2

  

      + (Pulser risetime)

2

 + (Oscilloscope risetime)

2

.

The risetime of the 419 is typically 3 ns. 

5.2.  FACTORY REPAIR

 
This instrument can be returned to ORTEC for
service and repair at a nominal cost. Our standard
procedure for repair ensures the same quality
control and checkout that are used for a new
instrument. Always contact the Customer Service
Department at ORTEC, (865) 482-4411, before
sending in an instrument for repair to obtain
shipping instructions and so that the required
Return Authorization Number can be assigned to
the unit. Write this number on the address label and
on the package to ensure prompt attention when it
reaches the factory. 

Summary of Contents for 142PC

Page 1: ...Model 142PC Preamplifier Operating and Service Manual Printed in U S A ORTEC Part No 717600 1202 Manual Revision B...

Page 2: ...n so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of t...

Page 3: ...2 2 5 ELECTRICAL AND MECHANICAL 2 3 INSTALLATION 3 3 1 CONNECTION TO PROPORTIONAL COUNTER 3 3 2 CONNECTION TO AMPLIFIERS 3 3 3 INPUT POWER 3 3 4 TEST PULSE 3 3 5 DETECTOR BIAS INPUT 4 4 OPERATION 4 4...

Page 4: ...iv...

Page 5: ...e Input connector on the preamplifier 2 Discharge the bias circuitry before making any connections to the Input connector and before disconnecting the preamplifier from the proportional counter 3 To d...

Page 6: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Page 7: ...the instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Page 8: ...viii...

Page 9: ...y long cable lengths can be used between the proportional counter and the preamplifier without disrupting the system stability An internal risetime compensation adjustment is accessible through a hole...

Page 10: ...rtional counter BIAS Accepts the bias voltage for the proportional counter from a bias supply TEST PULSE Accepts input voltage pulses from a pulse generator for instrument and system check and calibra...

Page 11: ...ther the E or T output connector is not being used it should simply be left open circuited unterminated 3 3 INPUT POWER Power for the 142PC is supplied through the Power Cable that is captive through...

Page 12: ...should be furnished with the bias supply module 4 OPERATION 4 1 GENERAL When the 142PC is installed according to the appropriate information in Section 3 it operates at all times when power is applie...

Page 13: ...The energy output signal from the preamplifier is a fast risetime voltage step with an exponential return to the baseline in about 75 s The polarity of signals through the E output is inverted from t...

Page 14: ...timing experiments typical risetimes of 25 to 150 ns for detector capacities from 1 to 100 pF the total cable length should be kept as short as possible even though it is terminated Due to vagaries i...

Page 15: ...ue only to the noise contribution of the preamplifier and main amplifier The analyzer can be calibrated in terms of keV per channel by observing two different pulser peaks of known energy and the FWHM...

Page 16: ...hed charge termination to measure the rise time of the 142PC through the T or E output connector Connect the 419 output through the charge terminator to the input of the 142PC and use an oscilloscope...

Reviews: