background image

4

3.  INSTALLATION

3.1.  CONNECTION TO DETECTOR

A direct connection with 93 

S

 or 100 

S

 shielded

cable should be made between the detector and the
Input connector on the preamplifier. For best
results, the length of this cable must be as short as
possible. This will not only minimize the preamplifier
noise (due to the capacitive loading of the cable)
but will also maintain the stability of the
preamplifier. The complex impedance presented to
the preamplifier input that is due to transmission
line effects acting on the detector system
impedance can disrupt the stability of the whole
system.  The interconnecting cable, which acts as
an impedance transformer, must be kept as short as
possible for the system to remain compatible with
the wide bandwidth of the preamplifier. Due to
vagaries in the detector system, a definite
maximum length cannot be specified but is typically
24 inches for the 142A and 15 inches for the 142B
and 142C. 

Type RG-62/U cable is recommended for the
detector to preamplifier connection. This is 93 

S

cable with a capacity of 13.5 pF/ft. 

After the input cable has been installed, the
electronic noise performance of the preamplifier can
be predicted by adding the capacity furnished by the
detector to the capacity of the cable. The cable
capacity can be calculated from its length and its
rated capacity per foot. Figures 2.1 and 2.2 show
typical performance at two commonly-used
amplifier shaping time constants, based on the total
input capacitance. 

3.2.  ENERGY OUTPUT CONNECTION TO

MAIN SHAPING AMPLIFIER

 
The E output of the preamplifier can be used to
drive a long 93 

S

  Iine to a shaping main amplifier

and is designed to be directly compatible with
ORTEC main amplifiers. It can be used with any
shaping main amplifier if a power supply is also
used to furnish the preamplifier power requirements
that are available on all ORTEC main amplifiers. 

3.3.  TIMING OUTPUT CONNECTION TO

TIMING MODULES

 
The T output of the preamplifier can be used to
drive a long, terminated 50 

S

 cable to a timing

module. A typical timing module is an amplifier, fast
discriminator, or a time-to-amplitude converter.
When not being used, the T output should be
terminated in 50 

S

For a positive detector bias voltage polarity, the T
output signal polarity is negative, since the timing
channel operates noninverting with respect to the
detector output. For ORTEC ruggedized surface-
barrier detectors which require a negative detector
bias polarity, the E output may be used as the
source of a negative timing pulse or the T output
can be inverted through a suitable amplifier.

3.4.  INPUT OPERATING POWER

 
Power for the 142 Preamplifiers is supplied through
the captive power cord and 9-pin Amphenol
connector. This connector can be attached to the
mating power connector on any ORTEC main
amplifier or 114 Preamplifier Power Supply. The
preamplifier's power requirements are added to the
operating power, requirements of the amplifier or
power supply to which it is connected. 

3.5.  TEST PULSE

 
A voltage test pulse for energy calibration can be
accepted through the Test input connector on the
142 without the use of an external terminator. The
Test input of the preamplifiers has an input
impedance of 93 

S

 and its circuitry provides charge

injection to the preamplifier input. The shape of this
pulse should be a fast rise time (less than 10 ns)
followed by a slow exponential decay back to the
baseline (200 to 400 

:

s). While test pulses are

being furnished to the Test input, connect either the
detector (with bias applied) or its equivalent
capacitance to the Input connector on the 142. 

The Test input may be used in conjunction with a
pulser such as the ORTEC 419 or 448 to calibrate
the preamplifier E Output amplitude in terms of

Summary of Contents for 142A

Page 1: ...Models 142A 142B and 142C Preamplifiers Operating and Service Manual Printed in U S A ORTEC Part No 627990 1202 Manual Revision D...

Page 2: ...n so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of t...

Page 3: ...ON 4 3 1 CONNECTION TO DETECTOR 4 3 2 ENERGY OUTPUT CONNECTION TO MAIN SHAPING AMPLIFIER 4 3 3 TIMING OUTPUT CONNECTION TO TIMING MODULES 4 3 4 INPUT OPERATING POWER 4 3 5 TEST PULSE 4 3 6 DETECTOR BI...

Page 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Page 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Page 6: ...e device to the Detector Input connector on the preamplifier 2 Discharge the detector bias circuitry before making ANY connections to the Detector Input connectorand before disconnecting the preamplif...

Page 7: ...ults however the T output should be terminated in 50 S when not in use A bias circuit is included to accept the operating voltage required by the surface barrier detector The bias input circuit in the...

Page 8: ...42B Nominally 20 mV MeV 142C Nominally 20 mV MeV ENERGY RANGE 142A 0 200 MeV 142B 0 400 MeV 142C 0 400 MeV RISE TIME 0 to 0 5 V Pulse at E output on 93S Load Fig 2 3 142A 5 ns at 0 pF 12 ns at 100 pF...

Page 9: ...s a differentiated output signal compatible with typical 50 S timing system requirements polarity is the same as the input pulse polarity Fig 2 4 2 4 CONNECTORS INPUT TEST E AND T BNC UG 1094 U BIAS S...

Page 10: ...y is also used to furnish the preamplifier power requirements that are available on all ORTEC main amplifiers 3 3 TIMING OUTPUT CONNECTION TO TIMING MODULES The T output of the preamplifier can be use...

Page 11: ...leakage current is small R3 should not be installed in the circuit because it would tend to degrade the noise performance 4 2 DETECTOR BIAS The amount of bias required by the detector is specified in...

Page 12: ...from the signal polarity at the detector output When the normal positive bias polarity is used for the detector the detector output pulses are negative and the E output of the preamplifier is positive...

Page 13: ...the 142C the adjustment has been made for 400 pF For optimum results for other input capacities the control should be adjusted under actual operating conditions If the control has been adjusted for op...

Page 14: ...alent energy The timing outputs should have the same polarity as the inputs with a scale factor of about 20 less than the signals through the E outputs 3 The noise contribution of the preamplifier may...

Page 15: ...eV 7 The noise performance of the preamplifier as measured by these methods should not differ significantly from that given in the specifications in Section 2 8 lf during testing of the preamplifier a...

Page 16: ...AIR This instrument can be returned to ORTEC for service and repair at a nominal cost Our standard procedure for repair ensures the same quality control and checkout that are used for a new instrument...

Reviews: