78
The ARTROMOT
®
-S3 device is subject
to particular precautions regarding
electromagnetic compatibility (EMC).
The device must be installed and put
into service strictly in compliance with
the EMC directives put forth in the
accompanying documents.
Portable and mobile RF communication
systems may affect the
ARTROMOT
®
-S3 device.
The ARTROMOT
®
-S3 device should not
be used adjacent to or stacked with
other equipment. If adjacent or stacked
use is necessary, ARTROMOT
®
-S3
should be observed to verify normal
operation in the configuration in which it
will be used.
If you detect damage or malfunctions
that may impair the safety of the patient
or of the operator, have the device
repaired before using it.
If it is necessary to replace assemblies
or cables, only the manufacturer's
original parts must be used to ensure
continued compliance with EMC
requirements after repair. This require-
ment applies to the power supply unit,
cables and cable lengths, drive unit
consisting of the motor and the control
system, the programming unit incl. the
coiled cable and the connector.
The equipment name ARTROMOT
®
-S3
used below always designates all
device models such as
ARTROMOT
®
-S3 and ARTROMOT
®
-S3
comfort.
9. IEC 60601-1-2:2001
9.1 Electromagnetic emissions
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
ARTROMOT
®
-S3 is intended for use in the electromagnetic environment specified below. It is the
responsibility of the customer or user to ensure that the ARTROMOT
®
-S3 device is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions to CISPR 11
Group 1
ARTROMOT
®
-S3 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions to CISPR 11
Class B
ARTROMOT
®
-S3 is suitable for use in all
establishments, including domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions to IEC
61000-3-2
not applicable
Voltage fluctuations/flicker
emissions to
IEC 61000-3-3
not applicable