English
43
The product described in this operation
manual must not be dispose of with
unsorted household or municipal waste.
It requires separate disposal. Please
contact ORMED.DJO for information
about the possible recycling of the
product.
Input ratings:
100 – 240 V AC/
50 – 60 Hz
Current
consumption:
850 – 370 mA
Fuses:
2 x T1A
Protection class:
I
Applied part:
type B
Max. load
on carriage:
20 kg
Dimensions
length:
96 cm
width:
35 cm
height:
min. 23 cm to
max. 56 cm
Adjustment ranges (min./max.):
Femur range:
approx. 31 – 49 cm
Lower leg range:
approx. 25 – 57 cm
Weight:
11 kg
Materials used:
ABS, POM
(Delrin 100),
PUR, PA, FR4,
aluminium, stainless
steel, brass
MDD
class 2a
Standards compliance:
IEC 60601-1:1988
+ A1:1991
+ A2:1995
Certification:
ANSI/UL 60601-1
CAN/CSA C22.2
No. 601.1
7. Environmental
protection statement
8. Specifications
Follow these steps to prepare the
ARTROMOT
®
-K1 classic for transport:
1. Adjust the femur length to 49 cm
and the tibia length to 42 cm.
2. Set the carriage to 0° by pressing
the START/STOP key when the
carriage is in the 0° position. This
will stop the carriage in that position
3. Push the power switch to turn off
the ARTROMOT
®
-K1 classic.
4. Disconnect the power cord and the
programming unit.
5. The device must be stored in its
original shipping box for transport.
ORMED GmbH cannot be held liable
for damage in transit, if the original
shipping box was not used.
6. Set the footplate to a horizontal
position.
7. Now slide the polystyrene pads
onto the ARTROMOT
®
-K1classic.
8. Place the power cord at the
bottom of the box before inserting
the ARTROMOT
®
-K1 classic including
the polystyrene pads.
9. Put the programming unit (19) in the
supplied box, and store both in the
ARTROMOT
®
-K1 classic box.
6.3 Transport
Programming Unit
Power Cord