background image

Imported by 

AIC International, 44980 Ste-Luce-sur-Loire, France 

PORTABLE INDOOR AIR QUALITY TESTER 

 

User manual 

 

Please read the instructions below carefully before use and keep this manual for future reference 

 

The portable air quality tester (product reference 23660) has been designed for indoor use only. 

 

The tester continuously measures the level of various pollutants and provides a practical, ever-ready, 
alternative prevention solution to comply with national regulations (e.g., French decree 2015-1000 of 
17th August, 2015: 2012-14) on the measurement of pollutants in the indoor air of public buildings and 
venues. 
 
Its primary purpose is to indicate air quality in enclosed atmospheres. The tester cannot be considered 
a substitute for the indoor air quality monitoring methods described in your relevant legislation. 

 

Technical features: 

 

Measures  concentrations  of  airborne  pollutant:  formaldehyde  (HCHO),  particulates  (PM1.0, 
PM2.5, PM10), and total volatile organic compounds (TVOC) in a room and displays them on 
the LCD display 

 

Measures and displays the room’s relative humidity and ambient temperature. 

 

Power supply: lithium battery 1200 mAh rechargeable via USB cable (included). 

 

Storage and carrying pouch. 

 

Detection and 

measurement range 

Formaldehyde (HCHO) level 

0.001 to 1.999 mg/m³ 

 

Total volatile organic 
compounds (TVOC) 

0.001 to 9.999 mg/m³ 

 

PM1.0, PM2.5, PM10 levels 

0 to 999 µg/m³. 

 

Temperature 

–10°C to +50°C 

 

Humidity 

20% to 85% RH. 

 

 

 

Accuracy 

Formaldehyde (HCHO). 

± 20% 

 

TVOC 

± 20% 

 

Particulates 

± 15% 

 

Temperature 

± 1°C 

 

Humidity 

± 5% 

 

 

 

Granularity: 

Formaldehyde (HCHO) 

0.001 mg/m³ 

 

TVOC 

0.001 mg/m³ 

 

Particulates 

1 µg/m³ 

 

Temperature 

1°C 

 

Humidity 

1% RH 

 

Operating conditions:  temperature –10°C to +45°C, humidity 20% to 85%, avoid condensation. 
Storage conditions:  temperature –10°C to 60°C, humidity 20% to 85%, avoid condensation. 
 
Sensor technology: electrochemical for formaldehyde, semiconductor for TVOC, laser for particulates. 

 

Summary of Contents for 23660

Page 1: ...s Measures concentrations of airborne pollutant formaldehyde HCHO particulates PM1 0 PM2 5 PM10 and total volatile organic compounds TVOC in a room and displays them on the LCD display Measures and di...

Page 2: ...and adapt to the atmosphere A countdown of this time in seconds is shown on the TVOC display line 4 To recover from the effects of packaging and shipping the device must be placed outdoors in a dry p...

Page 3: ...measured data points on the graphs select Clean Power saving mode Off Time to adjust the delay before the screen switches to standby mode Never or after 30 60 or 90 minutes If a set time is chosen the...

Page 4: ...is switched on If no readings are displayed make sure the Clear Log option is not activated Clean Step 4 Measurements and comfort ratings 1 Particulates are airborne particles resulting from wood bur...

Page 5: ...iod 2 Move away from the source of pollution Leave the room Ventilate 1 Guide values for Formaldehyde and Benzene based on French decree No 2015 1000 of the 17 August 2015 2012 14 Long exposure throug...

Page 6: ...ach threshold is coupled to a specific exposure time If the measured values regularly exceed these limits a professional diagnosis is advisable to identify the source of the pollutants The AQI Air Qua...

Page 7: ...held liable for damages arising from improper use or maintenance of the product Keep the appliance out of children s reach Troubleshooting guide Fault symptoms Checks solutions The tester does not tu...

Reviews: