7.3. - CHANGE THE FUSES
If the switch is in the position “on” and the lamp is connected to the mains but does not work, then check the fuse following these
instructions:
1) Remove the No.
10
fuses box cover using a screw driver. The fuse will come out with it
2) If it is damaged, replace it with a spare (250VF2A) to the left of the network cabling or (250VF5A) on the right.
3) Replace the use box cover.
8. - TECHNICAL CHARACTERISTICS
Electrical connection
100-240V / 47-63Hz
Luminous intensity at 100 cm
65.000 lux (± 10%)
Illumination intensity adjustment
from
20.000
to
65.000 lux
Light source
LED
Colour temperature
4.500 K (±5%)
CRI
>95
Power
28W
Ø Light head
300 mm
Field diameter
d
10
260 mm
Field diameter
d
50
130 mm
Depth of field
(L1 + L2)
170 cm
(80 + 90cm)
Colour
RAL 9002
Approximate Lamp weight
15 Kg
Battery type
Lead- acid free sealed
maintenance.
Battery Capacity
12V 12Ah
MODELS:
IG-65
M
Moveable lamp
IG-65
W
Wall mount lamp
IG-65
C
Ceiling lamp
IG-65
DC
Ceiling lamp with double arm
IG-65
MB
Moveable lamp with battery equipment
9. - PLANS AND DRAWINGS
See page 10
The fuse on the right 250VF5A network cable is initially
extracted. To connect the emergency system is necessary to
insert the fuse into the place.
Fig. 5 – General dimensions