ORDISI IG-65MB Maintenance And User Manual Download Page 8

 

Página 8 de 11 

 

4.4.-  RECOMENDATIONS FOR USE 

 

Move the lamp gently, avoiding brusque movements. 

If the lamp is not in use do not leave it on, since apart from wasting electricity you will reduce the useful life of the bulb considerably 

and overheat the cupola unnecessarily.

 

 

5. -  EMERGENCY EQUIPMENT

 

 

The emergency equipment has a lead acid battery 12V 12Ah sealed and maintenance free which allows grid independence. Once 

inserted the fuse 250VF5A (No. 

10

) will see a check for the LEDs. The battery status can be checked using the display located on 

the front base No. 

12

 

1.  The blue LED (B) indicates that emergency equipment is connected to mains.   

2.  The red LED (R) indicates that the emergency equipment is low, considering that in 30 minutes the lamp will turn off. 

3.  The orange LED (O) indicates that the emergency equipment is at half load.  

4.  The green LED (G) indicates that the emergency equipment is nearly charged 

 

 

 

 

6.- CLEANING 

 

Before doing anything make sure the lamp is off. 

The  lamp  must  be  cleaned  regularly  with  a  cloth  dampened  with  water,  taking  care  that  water  does not reach the inside of the 

equipment via the ventilation grille. For Example; blood, disinfectants, etc… a little neutral soap can be used. 

Do not use products 

containing acetone in any case.

 

Dry the lamp well before use. If there is liquid inside you must wait until it has evaporated. 

 

The handle located on the cupola is sterilizable by steam media at a temperature of 121 ° C with an exposure time of 20 minutes. 

 

 

7. - MAINTENANCE AND UPKEEP 

 

7.1. - PERIODIC CHECKS 

 

Periodically, according to hospital procedures, the maintenance service should carry out the following checks: 

1)  Check the general working order of the lamp. 

2)  Check the mobility of the arm and cupola. See point 6.2. 

3)  Check the equipment is still earthed and is correctly connected. 

4)  Check its general state (of cleanliness, etc…). 

 

 

         7.2. – ADJUST OF THE ARM 

 

Check the mobility of the arm and copula. If it’s necessary adjust them, the optimum adjustment point is when the lamp remains 

stable without touching it and be moved with a gentle movement. Do not over adjust the brakes, this only prematurely exhaust the 

articulations and impede movement. Adjust by screws No. 

13

,

14

 or 

15

     

      

 

 

 

It is advisable to keep the equipment connected to mains for maximum 

durability and reliability of emergency equipment. 

Fig. 4a – Cupola and fork adjustment. 

Fig. 4b – Arm adjustment. 

Summary of Contents for IG-65MB

Page 1: ...Mom 44 Edici n 1 Fecha 2017 01 10 Tecnolog a 2 Nave 7 08780 Pallej Barcelona Spain 34 93 334 01 12 FAX 34 93 440 25 64 comercial ordisi com www ordisi com...

Page 2: ...ibuidor ORDISI m s pr ximo o directamente a nuestra Central ORDISI S A Tecnolog a 2 08780 Pallej Barcelona Espa a Tel 34 93 334 01 12 Fax 34 93 440 25 64 e mail comercial ordisi com 2 RESPONSABILIADES...

Page 3: ...NE CEI as como con el Reglamento de Baja Tensi n La base enchufe en donde se conecte el equipo debe ser apta para clavijas seg n norma UNE 20315 y o DIN 49441 con bornes de 4 8 mm y toma de tierra lat...

Page 4: ...ue sta no penetre al interior de la c pula Para las manchas persistentes p e sangre desinfectantes etc se puede utilizar un poco de jab n neutro No utilizar bajo ning n concepto productos que contenga...

Page 5: ...Fuente de luz LED Vida LEDS 50 000 h Temperatura de color 4 500K 5 CRI 95 Potencia 28 W de la c pula 300 mm campo de trabajo d10 260 mm campo de trabajo d50 130 mm Profundidad de campo L1 L2 170 cm 8...

Page 6: ...ur L33 38 please contact your nearest ORDISI distributor or contact us to our head office directly ORDISI S A Tecnolog a 2 08780 Pallej Barcelona Espa a Tel 34 93 334 01 12 Fax 34 93 440 25 64 e mail...

Page 7: ...ion of the room where you plan to use the equipment should be in accordance with the rules UNE CIS as well as the Low Voltage Regulation The plug socket where the equipment is going to be connected mu...

Page 8: ...de of the equipment via the ventilation grille For Example blood disinfectants etc a little neutral soap can be used Do not use products containing acetone in any case Dry the lamp well before use If...

Page 9: ...10 Illumination intensity adjustment from 20 000 to 65 000 lux Light source LED Colour temperature 4 500 K 5 CRI 95 Power 28W Light head 300 mm Field diameter d10 260 mm Field diameter d50 130 mm Dep...

Page 10: ...P gina 10 de 11 9 PLANOS Y ESQUEMAS PLANS AND DRAWINGS Fig 6 Despiece de la c pula Exploded view of the cupola...

Page 11: ...P gina 11 de 11 Fig 7 Despiece de brazos y base rodable Exploded view of arms and base...