PROFIBUS DP
33
-
The following Error-Codes are supported (defined in the GSD file):
Channel_Diag(16) = "Target: Temperature too high"
Channel_Diag(23)= "Sensing head: Temperature too high"
Channel_Diag(17) = "Target: Temperature too low"
Channel_Diag(24)= "Sensing head: cable defect"
Channel_Diag(18) = "Box: Temperature to low"
Channel_Diag(25)= "Sensing head: Short circuit"
Channel_Diag(19) = "Box: Temperature too high"
Channel_Diag(26)= "Error 10 Status
alarm“
Channel_Diag(20) = "Error Statusbyte 24"
Channel_Diag(21) = "Error Statusbyte 28"
Channel_Diag(22) = "Sensing head: Temperature to low"
4.11.4 Process alarm
The process alarm for Optris0A90 is set up as following:
Byte
Data
Description
0x00
0x07
Header
0x01
0x02
Process alarm
0x02
Plug in position number
0x00: Status alarm
0x01
– 0x03 : Module
0x03
0x00
Fix
0x04
Bit 0: exceed of maximum boundary value
0x05
Bit 0: below boundary value
0x06,0x07
Current value