Technical Data [MSPro]
Temperature range
-32...760°C (-20...1440 °F)
± 1% or ± 1°C (20...760°C)
± 1,5°C (19,9...0°C)
± 2,5°C (-0,1...-20°C)
± 3°C (-20,1...-32°C)
Repeatability
± 0,75% or ± 0,75°C (20...760°C)
Optical resolution
40:1/ 13mm spot size in
260mm
Resolution (display)
0,1°C (0,1°F)
Response time (95%)
300 ms
Ambient temperature
0...50°C
Storage temperature
-20...60°C (without battery)
Spectral range
8...14µm
Emissivity/ Gain
0,100...1,100 adjustable
Functions
MIN, MAX, HOLD, °C/°F, Offset
Alarm functions
Visual and acoustic HIGH- and LOW-
alarm
Laser
< 1mW laser class IIa, laser beam with
9mm offset
PC interface
USB
Software
OptrisConnect Report software
Data logger
for 20 values
Input
for t/c probe type K
Weight/ Dimensions
180g, 190x38x45 mm
Battery
9V alkaline battery
Battery life time
20h (laser and backlight on 50%)/ 40h
(laser and backlight off)
Relative humidity
10-95% RH, non condensing at ambient
temperature < 30°C
Accuracy
(at T
amb
= 23 ±5°C)
Troubleshooting
Error/ Code
Problem
Action
HHH
object temperature
above range limit
choose object within
measuring range
LLL
object temperature
below range limit
choose object within
measuring range
battery indicator
low battery
replace battery
no display
low battery
replace battery
low battery
replace battery
laser deactivated
activate laser
laser does not
work