With this function you can select if the
maximum, minimum or t/c probe value [MSPro]
will be shown permanently in the upper part of
the display. After a measurement (during the
HOLD mode) you can also recall the
respectively non shown values by pressing the
U
P
button.
The t/c probe value will be displayed only if a
probe is connected. During the HOLD mode this
value will also be frozen.
A
LARM FUNCTIONS
To activate/ deactivate please press the
U
P
or
D
OWN
button if the
H
or
L
is shown in the
display. The alarm function is activated as soon
as the display shows the sign
.
After pressing
M
ODE
again the alarm values can
be adjusted using the
U
P
and
D
OWN
button.
If the temperature exceeds the selected
High
value
an acoustic signal will appear and the
display color will change to
RED
.
If the temperature falls below the selected
Low
value
an acoustic signal will appear and the
display color will change to
BLUE
.
°C/
°F
S
ETTING
Selection of the temperature unit.
O
FFSET
With this function you can set a linear offset
(+/-) to the temperature reading. It allows a field
calibration of several units showing exactly the
same values.
Reset Function
The unit can be reset to the factory default
values by pressing the
M
ODE
and
U
P
button
simultaneously (during HOLD mode). The Data
logger [MSPro] will not be deleted by this
procedure.
optris MS – DE2011-04-A
30