OP930D Functionality

NEW

AN-105

805

.

987

.

1700

sales

@

optotest

.

com         

engineering

@

optotest

.

com

2

O

ptoTest Corp.

4750 Calle Quetzal        

Camarillo, CA 93012

Doc: AN-105 Rev.B    12

/

13

/

13

For duplex measurements using the OP930D, connect two reference cables to the corresponding source channel, 

and then connect the reference connectors to their corresponding OPM channel.

NEW

Insertion Loss Reference

Figure 2: A reference setup for duplex cable measurement

Figure 3: A duplex return loss reference setup

Cables terminated with a PC polished connector exhibit the inherent 4% reflection (~14dB). This is the ideal 

situation for taking the reference. A clean connector interface on both ends, the APC connection at the front panel 

of the instrument, and the PC interface of the reference cable are all necessary.

Reference return loss by leaving the PC endface of an APC-PC launch cable unmated. Navigate to the RL screen for 

the first channel by pressing the right arrow button twice from the OPM screen. With the launch cable connected 

to the front panel, press and hold Ref to begin a return loss reference. Repeat this process for the second set. For 

methods of testing APC-APC reference cables, please consult AN-114 RL Testing for APC Cables.

NEW

Return Loss Reference

dB

Dual IL/RL Test Set

dBm

Ref

OP930D

IL

1310

nm

dB

Source

0.00

 

1550 nm

1310 nm

Detector

dB

0.00

 

-2.42

 

-1.72

 

dBm

dBm

Dual IL/RL Test Set

dBm

Ref

OP930D

RL

1310

nm

dB

Source

14.00

 

1550 nm

1310 nm

Detector

01

 

CH

 

POS

 

3.3m

 

77.3  F

 

minimal length is 3 meters

PC

APC

APC

SAVer**

HPR*

*HPR High Performance Cable

** SAVer Cable

APC

APC

HPR*

HPR*

SAVer**

SAVer**

APC

PC
PC

APC

HPR*

PC

SAVer**

APC

APC