OP930D Functionality
NEW
AN-105
805
.
987
.
1700
sales
@
optotest
.
com
engineering
@
optotest
.
com
3
O
ptoTest Corp.
4750 Calle Quetzal
Camarillo, CA 93012
Doc: AN-105 Rev.B 12
/
13
/
13
Connect the cable(s) to be measured to the reference end of the reference cable and the other end of the cable(s)
under test into the optical port of the power meter.
NEW
Testing
Once a proper reference has been taken the instrument will continuously update the return loss value or insertion loss
value depending on what mode it is in. The OPM and IL screens will show the optical power reading and insertion loss
values, respectively, for both channels simultaneously. Each channel has its own return loss mode so that the distance
and the value can be displayed simultaneously.
Figure 5: Front panel display for duplex cable
dB
dBm
Ref
IL
1310
nm
dB
-0.18
dB
-0.10
-2.42
-1.72
dBm
dBm
dB
Dual IL/RL Test Set
dBm
Ref
OP930D
IL
1310
nm
dB
Source
-0.18
1550 nm
1310 nm
Detector
dB
-0.10
-2.42
-1.72
dBm
dBm
Figure 4: Testing the insertion loss and return loss of a duplex cable
APC
APC
PC
PC
APC
APC
SAVer**
SAVer**
HPR*
HPR*
DUT***
DUT***
*HPR High Performance Cable
** SAVer Cable