www.cnoec.com.cn Page 17 of 17 [email protected]
8-2
Vid eo troubl esh ootin g
Trouble
Cause
Remedy
1
、
Incision image appears in the
video view
The pole is not in correct
position
Draw it to the correct position
2
、
Dirt appears in the video view
Dirt on the specimen
Clean the specimen
Dirt on the surface of objective
Clean the surface
3
、
Image is not clear while the
focus changing
The image is not clear in the
high magnification
Readjust the high magnification
4
、
No image on the TV screen
The draw pole is not in correct
position
Draw it to the correct position
Objective cover is not open
Open it
TV is not on Video channel
Choose the correct one
5
、
No image on the Monitor
Connection is not correct
Reconnect the circuit
Objective cover is not open
Open it
The input signal does not accord
with the signal be chosen on the
Monitor
Choose the correct signal model
6
、
The software run slowly or
the window of the view does not
come out
12V DC power does not be
connect
Connect the 12V DC power
No input signal of A/D board
Reconnect
the
C-Video
or
S-Video signal
The input signal does not accord
with the signal which is chosen
in the driver of the A/D board
Choose the correct signal model
which match the input signal
7
、
The image is not correct on
the view window
The CCD model chosen in the
driver of the A/D board does not
accord with the real CCD
Choose the correct CCD model
We Remain Right To Change Technical Data Without Notice Due To Upgrade.