  8    Trouble Shooting                                                     

The  performance  of  the  microscope  can’t  be  made  fully  because  of  unfamiliar  using,  this  table  will  give  some 

advices. 

8-1

 

General troubleshooting

 

Trouble 

Cause 

Remedy 

1

Double images 

Interpupillary  distance  is  not 
correct 

Readjust it 

Diopter adjustment is not correct 

Readjust it 

Magnification  of  each  eyepiece 
is not the same size 

Mount  the  same  size 
eyepiece 

2

Dirt appears in the field of view 

Dirt on the specimen 

Clean the specimen 

Dirt on the surfaces of eyepiece 

Clean the surface 

3

Image is not clear   

Dirt  on  the  surfaces  of  the 
objectives 

Clean the objectives 

4

Image is not clear while the focus 

changing 

Diopter adjustment is not correct 

Readjust the diopter 

Focus is not correct 

Readjust the focus 

5

The focusing knob is not smooth 

The focusing knob is too tight 

Loosen  it  to  a  suitable 
position 

6

The image is obscure because of 

the  head  slipping  down  by  itself 
during observation 

The focusing knob is too loose 

Tighten  it  to  a  suitable 
position 

7

Incision image appears in the field 

of view or of the video view 

The  pole  is  not  in  correct 
position 

Pull  or  push  it  to  the 
correct position 

8

The  image  on  the  monitor  is  not 

clear  when  the  focusing  knob  is 
turned. 

The focus of video is not correct 

Readjust the focus of video 
to a correct position 

9

Eyes fell tired easily 

Diopter adjustment is not correct 

Adjust the diopter 

Brightness of light is not correct 

Adjust the brightness 

10

Bulb  does  not  work  when  the 

switch is on 

No power supply 

Check the connection with 
the power supply 

The  bulb  was  not  inserted 
correctly 

Insert it correctly 

Bulb is wrong 

Replace with a new one 

11

Bulb is burned out suddenly 

Use the wrong bulb 

Replace with a correct one 

The voltage is too high 

Control the voltage 
Eg: use voltage regulator 

12

Brightness is not enough 

Use a wrong bulb 

Replace with a correct one 

The voltage is too low 

Increase the input voltage 

13

The  bulb  flickers  or  the 

brightness is unstable 

The bulb will burn out soon 

Replace with a new one 

The  bulb  was  not  inserted 
correctly 

Insert it correctly 

 

 

Summary of Contents for A23.3645 Series

Page 1: ...www cnoec com cn Page 1 of 17 sale cnoec com A23 3645 Series Zoom Stereo Microscope Instruction Manual Please read this manual carefully before use microscope...

Page 2: ...you take the films out of the big camera 1 2 MAINTENANCE 1 All lenses must be kept clean Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens...

Page 3: ...com 2 Nomenclature 2 1 A23 3645 Eyepiece shield Eyepiece Diopter adjusting ring Head Illumination device Zoom control knob Focusing Lamp lock screw Glass stage Clip Base Top variable light control Fu...

Page 4: ...www cnoec com cn Page 4 of 17 sale cnoec com 2 2 A23 3645 Diopter adjusting ring Eyepiece shield Zoom control knob Eyepiece Head lock screw Head Focusing knob Glass stage Base Clip...

Page 5: ...www cnoec com cn Page 5 of 17 sale cnoec com 3 Assemble 3 1 A23 3645 Eyepiece shield Eyepiece Head Head lock screw Clips Glass stage...

Page 6: ...pecimen slide 1 Turn the zoom control knob to the maximum magnification 2 Turn the diopter adjusting rings to the zero 3 Observe the specimen through the right eyepiece and make the jkj image clear by...

Page 7: ...the pillar stand 4 9 Choose the optical system 1 You can alternate the binocular observation and video capture by pushing or pulling the pole You can attain binocular observation by pushing the pole i...

Page 8: ...out the pole and adjust the focusing knob and then the image will appear on the computer screen clearly 5 If no image or the image without color it may be because the model of the input signal does no...

Page 9: ...the observation and put the switch on OFF again after auto balance or the color of the image will be distortion Fig 12 4 20 Adjust the brightness of the bottom light 1 Turn the adjustable light knob...

Page 10: ...r Head WF10x 20mm A53 3645 B Trinocular Head WF10x 20mm A53 3645 T Eyepiece WF10x 20mm High eyepoint A51 3621 1020 WF10x 22mm High eyepoint A51 3621 1022 WF10x 23mm High eyepoint A51 3621 1023 WF15x 1...

Page 11: ...www cnoec com cn Page 11 of 17 sale cnoec com...

Page 12: ...www cnoec com cn Page 12 of 17 sale cnoec com 6 Stands Holder Arm...

Page 13: ...www cnoec com cn Page 13 of 17 sale cnoec com...

Page 14: ...75X 14 3 21X 10 7 67 5X 3 3 33 75X 6 7 101 25X 2 2 135X 1 7 20X 10 14X 14 3 7X 28 6 21X 9 5 28X 7 1 90X 2 2 45X 4 4 135X 1 5 180X 1 1 7 2 CCD specification CCD Model Specification NTSC PAL Pick up Dev...

Page 15: ...epiece mag X Auxiliary objective mag Field number of eyepiece Diameter of field of view mm Zoom mag X Auxiliary objective mag Photo adaptor mag Zoom mag Auxiliary objective mag Eyepiece mag TV adaptor...

Page 16: ...uring observation The focusing knob is too loose Tighten it to a suitable position 7 Incision image appears in the field of view or of the video view The pole is not in correct position Pull or push i...

Page 17: ...ose the correct one 5 No image on the Monitor Connection is not correct Reconnect the circuit Objective cover is not open Open it The input signal does not accord with the signal be chosen on the Moni...

Reviews: