                                                                                                    A23.0910     

 

www.optoedu.com

  

Page 8 of 14

 

[email protected]

 

 

 

Fig. 10 

 

 

Fig. 11 

 

 

Fig. 12 

2-2-7

 

Assemble the Clip 

1)

 

Use the clip to avoid the specimen moving. 

2)

 

Insert the clip

① 

into the two holes of the base, as the 

arrowhead direction shown in the figure. (See Fig. 10) 

 

2-2-8

 

Fix the Microscope Body Position 

1)

 

Loosen  the  lock  knob

,  and  rotate  the  microscope 

body and the focusing bracket group left and right. 

2)

 

Match the center of zoom body group with the center 

of worktable, then lock the lock knob

. (See Fig. 11) 

 

2-2-9

 

Connect the LED Light Source and Power Cord 

1)

 

Make  sure  the  main  switch  is  at  “O”  (OFF) 

position.(See Fig. 12) 

2)

 

Insert  one  end  of  the optical fiber

① 

into  the  inner 

hole

② 

of the power supply box to the bottom, and align 

the flat  surface with  the lock screw

,  then lock it with 

the lock screw

3)

 

Insert  the  other  end  of  the  optical  fiber

① 

into  the 

hole of the base

, then lock it with the lock screw. 

4)

 

Insert  one end of the power adapter

⑥ 

into the inner 

hole

⑤ 

of the power supply box. 

5)

 

Insert  one  end  of  the  power  cable 

⑦ 

into  the 

corresponding  hole  of  the  power  adapter

,  and  the  

other end into the power supply socket. 

 

★ 

The  power  supply  box  supports  wide  voltage  as 

100-240V. 

★ 

Don’t  use  strong  force  when  the  power  cord is 

bended or twisted, otherwise it will be damaged. 

★ 

Use the special wire supplied by  our  company. If 

it’s  lost  or  damaged,  choose  one  with  the  same 
specifications. 

★ 

Connect  the  power  cord  appropriately  to make 

sure the instrument is connected to ground. 

Summary of Contents for A23.0910

Page 1: ...cnoec com To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use your instrument A23 0910 Zoom Stereo Microscope Parallel Light 0...

Page 2: ...djust the Focusing Tension 9 3 4 Adjust the Diopter and Focus 10 3 5 Adjust the Interpupillary Distance 10 3 6 Adjust the Elevation Angle of Observation Head 10 3 7 Turn on off the Magnification Fixed...

Page 3: ...l power adapter is used Wide voltage range is supported as 100 240V AC 5 Use the special wire supplied by our company 2 Maintenance 1 Keep all the surface of the lens clean Wipe the lens gently with a...

Page 4: ...Pole Aperture Diaphragm Adjusting Rod Tension Adjusting Knob Main Objective Worktable 1 Components A23 0 Microscope Body Focusing Bracket Group Coarse Focusing Knob Clip Base Group Trinocular Head Ey...

Page 5: ...embling Scheme to describe how to assemble the components and the numbers denote the assembling order Before assembling make sure there is no dust dirt or other materials which will disturb it Assembl...

Page 6: ...the focusing bracket group on the wrong side the microscope will turn over as the right view of Fig 3 2 2 2 Assemble the Microscope Body 1 Remove the decorative cover of the focus bracket group and l...

Page 7: ...hes the bottom Lock the eyepiece by the lock screw See Fig 7 Fig 7 Fig 8 Install or Remove the Reticle 1 Unscrew the pressure ring from eyepiece see Fig 8 2 Clean the reticle then install it into the...

Page 8: ...the optical fiber into the inner hole of the power supply box to the bottom and align the flat surface with the lock screw then lock it with the lock screw 3 Insert the other end of the optical fiber...

Page 9: ...stronger 1 Adjusting the adjustment knob of the reflector See Fig 14 can change the reflector surface also can change the lighting range after reflected 2 The adjustment knob of the reflector can be m...

Page 10: ...knob to the maximum magnification if the image is not clear enough when observed with the right eyepiece repeat Step 3 5 as above 6 Then rotate the zoom knob to the minimum magnification turn the diop...

Page 11: ...to the end with an inner hexagon spanner See Fig 19 Every magnification value on the zoom knob has a ticking sound fixed position 2 To turn off the ticking sound magnification fixed positon fuction r...

Page 12: ...ead which can used for binocular observation See Fig 22 When the light path selecting pole is pulled to the outmost all the light will enter the trinocular head which can used for TV Photography Fig 2...

Page 13: ...up onto the base with a M3 sunk key make the flat side of polarizer group close to the pillar then put back the worktable See Fig 24 2 Loop the analyzer group on the outside diameter of main objective...

Page 14: ...ght field of view doesn t superpose with the left one Interpupillary distance is incorrect Adjust the interpupillary distance Diopter adjustment is incorrect Adjust the diopter The eyepiece for the ri...

Reviews: