Before Use 

A23.0910 

 

www.optoedu.com

  

Page 3 of 14

 

[email protected]

 

1. Operation Notice 

 
 
 
 
 
 
 
 
 
 

 

Fig. 1 

 

 

1.

 

As  the  microscope  is  a  high  precision  instrument, 

always  operate  it  with  care,  and  avoid  physical  shake 
during the operation. 
2.

 

Do  not  expose  the  microscope  in  the  sun  directly, 

either  not  in  the  high  temperature,  damp,  dust  or  acute 
shake.  Make  sure  the  worktable  is  flat,  horizontal  and 
firm enough. 
Following  environment  is  required  when  operating: 
Indoor temperature: 5

℃~

40

, Max relative humidity: 

80%. 
3.

 

When moving the microscope, use both hands to hold 

the position

① 

and  

② 

of the microscope. (See Fig. 1) 

★ 

It  will  damage  the  microscope  by  holding  the 

observation  head,  focusing  knob,  body  or  objectives 
when moving. 

4.

 

External power adapter is used. Wide voltage range is 

supported as 100~240V AC. 
5.

 

Use the special wire supplied by our company. 

 

 

2. Maintenance 

1.

 

Keep all  the surface of the  lens clean.  Wipe the lens  gently  with a soft  lens tissue or blow 
away  the  dust.  Carefully  wipe  off  oil  or  fingerprints  on  the  lens  surfaces  with  tissue 
moistened with a little of 3:7 mixture of alcohol and ether. 

2.

 

Do not use organic solution to wipe the surface of other components, especially the surface 
of plastic components. Please use the neutral detergent if necessary. 

3.

 

Never  disassemble  the  microscope  yourself,  otherwise  it  will  influence  its  function  or 
damage it. 

4.

 

After using, cover the microscope with a dust cover to prevent it from the dust, and store it  
in a place free from moisture to avoid rusting. 

5.

 

To keep  the performance of  the microscope, please  check it periodically.  For  more details, 
please contact the agents nearby. 

 

 

Summary of Contents for A23.0910

Page 1: ...cnoec com To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use your instrument A23 0910 Zoom Stereo Microscope Parallel Light 0...

Page 2: ...djust the Focusing Tension 9 3 4 Adjust the Diopter and Focus 10 3 5 Adjust the Interpupillary Distance 10 3 6 Adjust the Elevation Angle of Observation Head 10 3 7 Turn on off the Magnification Fixed...

Page 3: ...l power adapter is used Wide voltage range is supported as 100 240V AC 5 Use the special wire supplied by our company 2 Maintenance 1 Keep all the surface of the lens clean Wipe the lens gently with a...

Page 4: ...Pole Aperture Diaphragm Adjusting Rod Tension Adjusting Knob Main Objective Worktable 1 Components A23 0 Microscope Body Focusing Bracket Group Coarse Focusing Knob Clip Base Group Trinocular Head Ey...

Page 5: ...embling Scheme to describe how to assemble the components and the numbers denote the assembling order Before assembling make sure there is no dust dirt or other materials which will disturb it Assembl...

Page 6: ...the focusing bracket group on the wrong side the microscope will turn over as the right view of Fig 3 2 2 2 Assemble the Microscope Body 1 Remove the decorative cover of the focus bracket group and l...

Page 7: ...hes the bottom Lock the eyepiece by the lock screw See Fig 7 Fig 7 Fig 8 Install or Remove the Reticle 1 Unscrew the pressure ring from eyepiece see Fig 8 2 Clean the reticle then install it into the...

Page 8: ...the optical fiber into the inner hole of the power supply box to the bottom and align the flat surface with the lock screw then lock it with the lock screw 3 Insert the other end of the optical fiber...

Page 9: ...stronger 1 Adjusting the adjustment knob of the reflector See Fig 14 can change the reflector surface also can change the lighting range after reflected 2 The adjustment knob of the reflector can be m...

Page 10: ...knob to the maximum magnification if the image is not clear enough when observed with the right eyepiece repeat Step 3 5 as above 6 Then rotate the zoom knob to the minimum magnification turn the diop...

Page 11: ...to the end with an inner hexagon spanner See Fig 19 Every magnification value on the zoom knob has a ticking sound fixed position 2 To turn off the ticking sound magnification fixed positon fuction r...

Page 12: ...ead which can used for binocular observation See Fig 22 When the light path selecting pole is pulled to the outmost all the light will enter the trinocular head which can used for TV Photography Fig 2...

Page 13: ...up onto the base with a M3 sunk key make the flat side of polarizer group close to the pillar then put back the worktable See Fig 24 2 Loop the analyzer group on the outside diameter of main objective...

Page 14: ...ght field of view doesn t superpose with the left one Interpupillary distance is incorrect Adjust the interpupillary distance Diopter adjustment is incorrect Adjust the diopter The eyepiece for the ri...

Reviews: