background image

 

www.cnoec.com.cn                                                                          Page 6 of 10                                                                    [email protected] 

 

magnification.  In  order  to  observe  a  large  viewfield,  set  0.4X  low-magnification 
objective    to the    whorl (at the downside of objective) and then    renew focusing . 

 

 

 

 

 

 

Fig.4 

To  order  to  keep  both  left  and  right  objective  at  the  same  magnification,  the 

operator  must  make  focusing  by  the  following  methods  :  separately  place  the 

magnification reticles(30) on two table plates, and view them through eyepieces until 

two  reticles’  scale    can    completely  coincide  .  If  the  left  and  right  scales  are  not 

matched  ,  turn    the  magnification  correcting  knob  until  the  reticles’  images  are 

sharp , and then move the reticles to keep two    reticles’ scales completely coincide. 

Repeat several operations , you may make sure that both left and right objective are at 

the  same  magnification  .  When  change  the  objective’s  magnification,  make  some 

adjustment according to the above methods. 

6.Bullet holder : shown as Fig.5 

 
 
 
 
 
 

To observe breech-face markings                      To observe bullet profile 

 

 

 

 

 

To observe the bottom of cartridge                    To observe cutting trace 

Summary of Contents for A18.1828-LCD

Page 1: ...A18 1828 LCD LCD Comparison Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Page 2: ...tion With our success in solving the technical problem of the adjustability of its separation line width its quality has therefore reached the international advanced standard This model is an ideal in...

Page 3: ...Horizontal linkage knob 5 Working table lifting knob 6 LED light source 7 Objective 8 Six holes nosepiece 9 Separation line window 10 Separation line adjusting knob 11 Video recording tube 12 CCD opti...

Page 4: ...bes insert the required eyepieces into the tubes 3 Plug one end of the power line into the socket on the base the other to the socket with 110V 220V ground protection 4 After the above steps are compl...

Page 5: ...n line The separation line should be a fine and black straight one Turn the separation line adjusting knob 10 to move the separation line you may get the left and right single viewfield observation cu...

Page 6: ...m through eyepieces until two reticles scale can completely coincide If the left and right scales are not matched turn the magnification correcting knob until the reticles images are sharp and then mo...

Page 7: ...rvation draw out the lever for photographing Fit the back of camera into the adapter 13 and then insert to the instrument Observe the object through viewframe adjust both left and right stages to obta...

Page 8: ...enances The instrument must be kept in shade cool and dry place free of dust acid alkali and vapor As soon as work is finished the dustcover should be put on it The instrument must be kept clean The d...

Page 9: ...ective 0 5x 1x 1 5x 2x 4x 6x 2 pcs each 5W LED Light 2 pcs Auxiliary Lens 0 4X 2 0x 2 pcs each 3 0 M CCD Built in Main Body 1 pcs Polarizer Analyzer 2 pcs each Power cable 1 pcs Coaxial light device 2...

Page 10: ...www cnoec com cn Page 10 of 10 sale cnoec com...

Reviews: