4.2 Installation   
4.2.1 Please confirm the installation condition meets 4.1; 
4.2.2 Put out the main body and place it in table, and loose the main body 
fixing screw, then put the cost cover out; 
4.2.3 Put out the intermediate, fix it into the main body, then tighten the 
main  body  fixing  screw;  If  it  is  Transmitting  &  Reflecting  Polarizer 
Microscope, put out the reflecting illumination    unit to fix it.   
    Installation  of  reflecting  illumination  unit:  Put  out  the  halogen  lamp 
house, fix it onto the back of reflecting illumination unit by the fixed screw, 
connect  the  power  wire  of  halogen  lamp  house  into  the  power  output 
socket of power house.   
4.2.4 Fix the head into the intermediate; 
4.2.5 Insert the eyepieces into the tubes; 
4.2.6 Put out the dust cover of the nosepiece, and turn the coarse focusing 
knob to lower the stage, and find the objective hole with yellow mark in 
the nosepiece. Fix the 10X objective into the hole, and turn the nosepiece 
clockwise, fix the other objectives as per the power. 
4.2.7 Loose the condenser lock thumb and fix the condenser. 

 

5. Operation 

5.1  Turn  the  power  switch  and  the  potentiometer  to  adjust light to  be 
available; 
 
5.2 Fix the sample on the stage, and move it into the path; 
 
5.3 Turn the nosepiece to put 10X objective into light path, and turn the 
focus knobs to get clear image. 
 
5.4 Confirm the polarizing vibrancy direction 
The  polarizing  vibrancy  direction  has  set  to  be  at  west-east  in  factory 
when the scale of the polarizing is 0°. 
 
5.5 Check polarizing and analyzer 
The field should be dark completely (when there is no sample) when the 
scale  of  polarizer  and  analyzer  is  0°.  Please  check  the  position  of  the 

Summary of Contents for A15.2601-T

Page 1: ...15 2601 T Polarizing Microscope Transmit Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Page 2: ...Contents 1 Purpose 2 2 Specification 2 3 Parts Name 4 4 Installation 8 5 Operation 9 6 Maintenance 11 7 Troubleshooting 13 Outfits 14...

Page 3: ...parent non transparent mineral sample 2 Specification 2 1 Total Magnification 2 1 1 Transmitting Objective Eyepiece 4X 10X 20X 40X 60X 100X 10X 40X 100X 200X 400X 600X 1000X 2 1 2 Reflecting Objective...

Page 4: ...piece Quadplex or quintuple nosepiece center adjusting nosepiece spanner 2 8 Revolving Round Stage Diameter 174mm 360 part scale 6 per scale 2 9 Focusing System Coaxial coarse and fine focusing knobs...

Page 5: ...ary Inlet Analyze r Centering Nosepiece Condens Polarizer Field Diaphragm Revolver Aperture Diaphragm Centering Screw Condenser Thumb Screw Main Body Fixing Screw Stage Centering Screw Clamp Bertrand...

Page 6: ...wer Switch Coarse Focusing Tension Adjusting Ring Left Fine Focusing Knob E intermediate C mount Analyzer Screw Objectiv es Bertrand Lens Locking Screw cursores Adjustable level for swing out Condense...

Page 7: ...Power house Potentiometer Power switch Fuse Input socket Output socket Halogen lamp house Halogen Lamp Centering Knob Lamp house fix screw Ground glass slider Aperture diaphragm lever Field diaphragm...

Page 8: ...4 The microscope should be used in environment of indoor temperature 0 40 C and maximum relative humidity 85 4 1 5 Please pay attention to prevent microscope from violent shake and vibration in applic...

Page 9: ...4 2 6 Put out the dust cover of the nosepiece and turn the coarse focusing knob to lower the stage and find the objective hole with yellow mark in the nosepiece Fix the 10X objective into the hole an...

Page 10: ...high power objective Adjusting Details 5 9 1 After getting clear image choose one image point I at the observational position then moving mineral sample making I in cross division line s point and it...

Page 11: ...the mixture of alcohol and ether proportion 1 4 6 1 2 Alcohol and ether all are burnt early please take them away from fire Be careful for turn on and off power 6 1 3 Don t clean painted metal and gal...

Page 12: ...3 Lay aside the microscope reliably unscrew the knurled thumb screw of the lamp housing cover on the underside of base 6 3 4 Pull over the lamp housing cover 6 3 5 Pull out the bulb should be replaced...

Page 13: ...er glass 0 17mm Cover glass of specimen isn t in up direction Place specimen correctly Surface of objective lens isdirty especially it is easy for the front lens of 40X objective to dip in immersion o...