-8- 

polarizer and the analyzer if not so. 
 
5.6 Choose the complementary slip as per the sample, then insert into the 
slip. 
 
5.7 Put into the Bertrand Lens Slip in the condition of polarizing; 
5.8 Adjust the center of the stage. 
 
5.9 Adjustment for round stage rotating center 
The rotating axis of the round stage should be in the same line with the 
axis of optical system. When rotating the stage, the sample’s center would 
coincide with view field’s cross division line intersection point, and image 
should  move  around  cross  division  line  intersection  point  in  circular 
motion, image should always in the view field. If not, accurate data can’t 
be  got,  and  the  operation  can’t  be  done  correctly,  especially  for  high 
power objective. 
Adjusting Details: 
 
5.9.1  After  getting  clear  image,  choose  one  image  point  (

  I

  )  at  the 

observational position, then moving mineral sample , making 

in cross 

division line’s point, and it is the center of field (mark: 

O

). 

 

5.9.2 Stage must rotate at least one circle, if actual rotating center (mark: 

O

)  isn’t  coincided  with  field  center 

O

1  , 

image  point 

I

  will  do  circular 

motion around stage rotating center 

O

 . 

 
5.9.3 The angle from rotating stage to initial position is 180°, point 

I

 will 

move from 

O

1

 

to the full distance position , the symmetric point to point O. 

Adjusting  the stage focusing screw, point 

I

 will move  to point

  O

, then 

move mineral sample gently, make point I coincided with field center 

O

1 . 

 
5.9.4 Rotating stage one circle again, if image point 

isn’t off point

 O

1

,

 it 

means point

 O

 has coincided with point 

O

1

then the center adjustment is 

finished. If not, please repeat step2 , 3, 4. 
 
 
 

Summary of Contents for A15.2601-T

Page 1: ...15 2601 T Polarizing Microscope Transmit Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Page 2: ...Contents 1 Purpose 2 2 Specification 2 3 Parts Name 4 4 Installation 8 5 Operation 9 6 Maintenance 11 7 Troubleshooting 13 Outfits 14...

Page 3: ...parent non transparent mineral sample 2 Specification 2 1 Total Magnification 2 1 1 Transmitting Objective Eyepiece 4X 10X 20X 40X 60X 100X 10X 40X 100X 200X 400X 600X 1000X 2 1 2 Reflecting Objective...

Page 4: ...piece Quadplex or quintuple nosepiece center adjusting nosepiece spanner 2 8 Revolving Round Stage Diameter 174mm 360 part scale 6 per scale 2 9 Focusing System Coaxial coarse and fine focusing knobs...

Page 5: ...ary Inlet Analyze r Centering Nosepiece Condens Polarizer Field Diaphragm Revolver Aperture Diaphragm Centering Screw Condenser Thumb Screw Main Body Fixing Screw Stage Centering Screw Clamp Bertrand...

Page 6: ...wer Switch Coarse Focusing Tension Adjusting Ring Left Fine Focusing Knob E intermediate C mount Analyzer Screw Objectiv es Bertrand Lens Locking Screw cursores Adjustable level for swing out Condense...

Page 7: ...Power house Potentiometer Power switch Fuse Input socket Output socket Halogen lamp house Halogen Lamp Centering Knob Lamp house fix screw Ground glass slider Aperture diaphragm lever Field diaphragm...

Page 8: ...4 The microscope should be used in environment of indoor temperature 0 40 C and maximum relative humidity 85 4 1 5 Please pay attention to prevent microscope from violent shake and vibration in applic...

Page 9: ...4 2 6 Put out the dust cover of the nosepiece and turn the coarse focusing knob to lower the stage and find the objective hole with yellow mark in the nosepiece Fix the 10X objective into the hole an...

Page 10: ...high power objective Adjusting Details 5 9 1 After getting clear image choose one image point I at the observational position then moving mineral sample making I in cross division line s point and it...

Page 11: ...the mixture of alcohol and ether proportion 1 4 6 1 2 Alcohol and ether all are burnt early please take them away from fire Be careful for turn on and off power 6 1 3 Don t clean painted metal and gal...

Page 12: ...3 Lay aside the microscope reliably unscrew the knurled thumb screw of the lamp housing cover on the underside of base 6 3 4 Pull over the lamp housing cover 6 3 5 Pull out the bulb should be replaced...

Page 13: ...er glass 0 17mm Cover glass of specimen isn t in up direction Place specimen correctly Surface of objective lens isdirty especially it is easy for the front lens of 40X objective to dip in immersion o...

Reviews: