7. Troubleshooting 

In the period of using BK series microscope, if there is any trouble occurs, please 

referring to the following sheet listed some common troubleshooting resolve them. 

Trouble 

Causation 

Remedy 

Switch on but bulb dark

 

Plug is unreliable 

Plug in again 

Bulb is broken 

Change bulb 

Fuse is broken 

Change fuse 

Bulb is flickering or 
brightness is unsteady 

Bulb is unstable 

Insert it again 

Bulb is broken 

Replacing bulb 

Brightness of view field 
isn’t enough or is 
Uneven 

Bulb specification doesn’t   
meet the requirement 

Replacing bulb 

Brightness isn’t adjusted 
correctly 

Adjust rotation 
potentiometer 

Objective isn’t in correct 
position 

Make the objective in correct 
position 

The size of iris aperture is too 
small 

Adjust the size of iris aperture 

Brightness of view field 
isn’t enough or is 
Uneven 

Lens (objective,eyepiece, 
condenser, light collector) has 
dust 

Clean it 

Position of condenser is too low  Higher condenser 

Image isn’t clear 
  (contrast or definition 
isn’t enough) 

Cover glass of specimen doesn’t 
meet the 
requirement 

Use required thickness cover 
glass (0.17mm) 

Cover glass of specimen isn’t in 
up direction 

Place specimen correctly 

Surface of objective lens 
isdirty (especially it is easy for 
the front lens of 40X objective 
to dip in immersion oil) 

Clean it 

Immersion oil isn’t used for 
100X objective (oil) 

Use immersion oil 

Immersion oil doesn’t meet the 
requirement 

Use immersion oil 
supplied by us 

There is bubble in immersion oil  Clear the bubble way 
Size of iris aperture isn’t   
proper 

Adjust the size of iris 
aperture 

Position of condenser is too low 

Readjust the position of 
condenser 

One side of image is 
dark or image is 
moving as focusing   

Objective isn’t in correct 
position 

Make the objective in 
correct position 

 

Specimen isn’t placed correctly 

Place specimen levelly on 
stage and clip it with clamp 

Objective touches 
specimen as changing 
low times objective to 
high times objective 

Cover glass of specimen isn’t in 
up direction 

Place specimen correctly 

 

Cover glass doesn’t meet the 
requirement 

Use required thickness 
cover glass (0.17mm) 

Image observed by two 
eyes aren’t in 
superposition entirely. 

Interpupilary distance isn’t 
adjusted correctly   

Adjust interpupilary   
distance according to two 
eyes 

It is easy for eyes to be 
tired during observing   

Diopter isn’t adjusted correctly  Readjust diopter 

 

Summary of Contents for A15.2601-T

Page 1: ...15 2601 T Polarizing Microscope Transmit Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Page 2: ...Contents 1 Purpose 2 2 Specification 2 3 Parts Name 4 4 Installation 8 5 Operation 9 6 Maintenance 11 7 Troubleshooting 13 Outfits 14...

Page 3: ...parent non transparent mineral sample 2 Specification 2 1 Total Magnification 2 1 1 Transmitting Objective Eyepiece 4X 10X 20X 40X 60X 100X 10X 40X 100X 200X 400X 600X 1000X 2 1 2 Reflecting Objective...

Page 4: ...piece Quadplex or quintuple nosepiece center adjusting nosepiece spanner 2 8 Revolving Round Stage Diameter 174mm 360 part scale 6 per scale 2 9 Focusing System Coaxial coarse and fine focusing knobs...

Page 5: ...ary Inlet Analyze r Centering Nosepiece Condens Polarizer Field Diaphragm Revolver Aperture Diaphragm Centering Screw Condenser Thumb Screw Main Body Fixing Screw Stage Centering Screw Clamp Bertrand...

Page 6: ...wer Switch Coarse Focusing Tension Adjusting Ring Left Fine Focusing Knob E intermediate C mount Analyzer Screw Objectiv es Bertrand Lens Locking Screw cursores Adjustable level for swing out Condense...

Page 7: ...Power house Potentiometer Power switch Fuse Input socket Output socket Halogen lamp house Halogen Lamp Centering Knob Lamp house fix screw Ground glass slider Aperture diaphragm lever Field diaphragm...

Page 8: ...4 The microscope should be used in environment of indoor temperature 0 40 C and maximum relative humidity 85 4 1 5 Please pay attention to prevent microscope from violent shake and vibration in applic...

Page 9: ...4 2 6 Put out the dust cover of the nosepiece and turn the coarse focusing knob to lower the stage and find the objective hole with yellow mark in the nosepiece Fix the 10X objective into the hole an...

Page 10: ...high power objective Adjusting Details 5 9 1 After getting clear image choose one image point I at the observational position then moving mineral sample making I in cross division line s point and it...

Page 11: ...the mixture of alcohol and ether proportion 1 4 6 1 2 Alcohol and ether all are burnt early please take them away from fire Be careful for turn on and off power 6 1 3 Don t clean painted metal and gal...

Page 12: ...3 Lay aside the microscope reliably unscrew the knurled thumb screw of the lamp housing cover on the underside of base 6 3 4 Pull over the lamp housing cover 6 3 5 Pull out the bulb should be replaced...

Page 13: ...er glass 0 17mm Cover glass of specimen isn t in up direction Place specimen correctly Surface of objective lens isdirty especially it is easy for the front lens of 40X objective to dip in immersion o...

Reviews: