Before Use 

                                                                   

A15.0908 Series 

 

1. Operation Notice

 

1. As the microscope is a high precision 
instrument, always operate it with care, and 
avoid physical shake during the operation. 
2. Do not expose the microscope in the sun 
directly, either not in the high temperature, damp, 
dust or acute shake. Make sure the worktable is 
flat and horizontal. Following environment is 
required when operating: Indoor temperature: 5

℃~

40

, Max relative humidity: 80%. 

3. When moving the microscope, use one hand to 
hold its front base and the other hand to hold its 
arm, and lay it down carefully (See the Fig. left). 

  It will damage the microscope by holding 

the stage, focusing knob, head or light source 
when moving. 

4. When working, the surface of light source will 
be very hot. Make sure there is enough room for 
the heat dissipating around the light source. 
5. Connect the microscope to the ground to avoid 
lightning strike. 
6. For safety, make sure the power switch is at 
“O” (OFF) and power it off before replacing the 
bulb or fuse (See the Fig. left), and wait until 
both the bulb and bulb holder have cooled down. 

Bulb selected only: 12V/50W Halogen bulb. 

7. Wide voltage range is supported as 100~240V. 
Additional transformer is not necessary. Make 
sure the voltage is in this range. 
8. Use the special wire supplied by our company. 

2. Maintenance 

(1) Wipe the lens gently with a soft tissue. Carefully wipe off the oil marks and fingerprints on 
the lens surfaces with a tissue moistened with a small amount of 3:7 mixture of alcohol and 
ether or dimethylbenzene. 

  As the alcohol and ether is flammable, don’t place these chemical near to fire or fire 

source. For example, when turning on/ off the electrical device, please use these chemical in 
a ventilated place. 

(2) Don’t use organic solution to wipe the surfaces of other components. Please use the neutral 
detergent if necessary. 
(3) If the microscope is damped by liquid when using, power it off immediately and wipe it dry. 
(4) Never disassemble the microscope, otherwise the performance will be affected or the 
instrument will be damaged. 
(5) After using, cover the microscope with a dust cover. 

3. Safety Sign 

Sign Signification 

 

 

The surface gets hot and don’t touch it with bare hand.

 

 

Study the instructions before use. Unsuitable operation would lead to 
person hurt or instrument faulty. 

Main switch ON 

O Main 

switch 

OFF 

 

Portable 
Position 

Handhold  
Position 

Main  
Switch 

Summary of Contents for A15 0908 Series

Page 1: ...and maintenance to the A15 0908 Series Polarizing Microscope Please study this manual thoroughly before operating and keep it with the instrument The manufacturer reserves the rights to the modificat...

Page 2: ...e cap 10 3 7 Adjust the Objective Center 10 3 8 Center the Condenser 11 3 9 Adjust the Field Diaphragm 12 3 10 Adjust the Aperture Diaphragm 12 3 11 Adjust the Analyzer 13 3 12 Adjust the Polarizer 13...

Page 3: ...ed only 12V 50W Halogen bulb 7 Wide voltage range is supported as 100 240V Additional transformer is not necessary Make sure the voltage is in this range 8 Use the special wire supplied by our company...

Page 4: ...er Adjustment Ring Epi Illuminator Group Upper Light Source Group Middle Head Bertrand Lens Centering Screw Objective Nosepiece Light Adjusting Knob Objective Body Coarse Adjusting Limit Knob Fine Foc...

Page 5: ...ss Flashboard Group Polarizer Flashboard Group Stage Lock Screw Tension Adjustment Ring Stage Clip TV and Photography Device Lock Screw Light Path Selecting Pole Trinocular Head Condenser Center Adjus...

Page 6: ...es 2 1 Assembling Scheme Following is the Assembling Scheme and the numbers denote the assembling order Before assembling make sure there is no dust or dirt Assemble carefully and do not scrap any par...

Page 7: ...ert the pins into the receptacles Make sure the bulb is vertical See Fig 3 3 Loosen the bulb lock sheet to lock the bulb and put on the cover group Before replacing cut off the main power supply and w...

Page 8: ...ser into the innermost of base 5 Screw down the condenser lock screw and raise the condenser to the highest position with the condenser up down knob 2 2 4 Assemble the Stage Clip Plug the two stage cl...

Page 9: ...ddle head into the hole on the Epi Illuminator with a little left inclined and then screw down the lock screw The screw head should be pushed into the V slot on the right side of the middle head coatt...

Page 10: ...on plug into the aviation socket of the microscope 4 Insert one end of the power cord into the microscope power socket 5 Insert the other end of the power cord into the power supply socket Cancel Step...

Page 11: ...t the switch to and push the spectroscope pole to the innermost to move the spectroscope into the light path then adjust the light adjustment knob until the illumination is comfortable for observation...

Page 12: ...cusing knob for clear details 3 3 Adjust the Focusing Tension If the handle is very heavy when coarse focusing or the specimen leaves the focus plane after focusing or the stage declines itself please...

Page 13: ...avoid damaging to both glasses and eyepieces 2 Open the eye cap if the user doesn t wear glasses so that it can prevent stray light disturbing the observation See Fig 20 3 7 Adjust the Objective Cente...

Page 14: ...g 23 Move the front lens of the condenser into light path when use the objective beyond 20X 3 Rotate the 20X objective into the light path and focus the specimen 4 Rotate the field diaphragm adjustmen...

Page 15: ...d of view then the field diaphragm is centered correctly See Fig 26 5 In actual use enlarge the field diaphragm a little to make its image circumscribed to the field of view For Transmitted Illuminati...

Page 16: ...the analyzer is shifted into the light path See Fig 29 2 Loosen the lock screw turn the analyzer turntable to change the vibrated direction of the analyzer A grid value of scale is 2 and the adjustin...

Page 17: ...an Bertrand lens and make the vibration direction of polarizer be vertical with the analyzer 3 14 Use the Compensator The compensator is inserted into the slot of the middle head in 45 declined The su...

Page 18: ...a then turn the coarse and fine focusing knobs until the image is clear 3 Rotate the Bertrand lens switch turntable locate it on position B to move the Bertrand lens into light path At that time the b...

Page 19: ...nermost for binocular observation pull the light path selecting pole to the outermost for trinocular observation TV and photography See Fig 35 3 19 2 Assemble and Use the TV Device 1 Loosen the lock s...

Page 20: ...pole to the outmost then operate the photography device according to its instructions 3 20 Use the Digital Viewing Head 3 20 1 Connect the Power Cord Insert the output connector of the transformer in...

Page 21: ...sabled when the white balance switch is popup while it is enabled when the switch is pressed down The switch is mainly for adjusting the color of the microscope TV image See Fig 40 When using the digi...

Page 22: ...k or not even Stain or dust has accumulated on the lens objectives or eyepieces Clean the lens Stains have accumulated on the specimen Clean the specimen 4 Stain or dust is observed in the view field...

Page 23: ...slide is not fixed correctly Adjust it correctly 5 Cannot move the slide smoothly The movable specimen holder is not fixed properly Fix it correctly 3 Electrical System No power supply Check the conn...

Page 24: ......