background image

 

 

Version:V1.0

 

 

   
 
 

A15.0908 Series 
Polarizing Microscope 
INSTRUCTIONS

 

 

 

 
 
 
 
 
 
 
 
 
 
 

This instruction manual is for the operation guide, troubleshooting and 
maintenance to the A15.0908 Series Polarizing Microscope. Please study this 
manual thoroughly before operating, and keep it with the instrument. The 
manufacturer reserves the rights to the modifications by technology 
development. On the basis of operation ensured, technical specifications 
may be subject to changes without notice.

 

Summary of Contents for A15 0908 Series

Page 1: ...and maintenance to the A15 0908 Series Polarizing Microscope Please study this manual thoroughly before operating and keep it with the instrument The manufacturer reserves the rights to the modificat...

Page 2: ...e cap 10 3 7 Adjust the Objective Center 10 3 8 Center the Condenser 11 3 9 Adjust the Field Diaphragm 12 3 10 Adjust the Aperture Diaphragm 12 3 11 Adjust the Analyzer 13 3 12 Adjust the Polarizer 13...

Page 3: ...ed only 12V 50W Halogen bulb 7 Wide voltage range is supported as 100 240V Additional transformer is not necessary Make sure the voltage is in this range 8 Use the special wire supplied by our company...

Page 4: ...er Adjustment Ring Epi Illuminator Group Upper Light Source Group Middle Head Bertrand Lens Centering Screw Objective Nosepiece Light Adjusting Knob Objective Body Coarse Adjusting Limit Knob Fine Foc...

Page 5: ...ss Flashboard Group Polarizer Flashboard Group Stage Lock Screw Tension Adjustment Ring Stage Clip TV and Photography Device Lock Screw Light Path Selecting Pole Trinocular Head Condenser Center Adjus...

Page 6: ...es 2 1 Assembling Scheme Following is the Assembling Scheme and the numbers denote the assembling order Before assembling make sure there is no dust or dirt Assemble carefully and do not scrap any par...

Page 7: ...ert the pins into the receptacles Make sure the bulb is vertical See Fig 3 3 Loosen the bulb lock sheet to lock the bulb and put on the cover group Before replacing cut off the main power supply and w...

Page 8: ...ser into the innermost of base 5 Screw down the condenser lock screw and raise the condenser to the highest position with the condenser up down knob 2 2 4 Assemble the Stage Clip Plug the two stage cl...

Page 9: ...ddle head into the hole on the Epi Illuminator with a little left inclined and then screw down the lock screw The screw head should be pushed into the V slot on the right side of the middle head coatt...

Page 10: ...on plug into the aviation socket of the microscope 4 Insert one end of the power cord into the microscope power socket 5 Insert the other end of the power cord into the power supply socket Cancel Step...

Page 11: ...t the switch to and push the spectroscope pole to the innermost to move the spectroscope into the light path then adjust the light adjustment knob until the illumination is comfortable for observation...

Page 12: ...cusing knob for clear details 3 3 Adjust the Focusing Tension If the handle is very heavy when coarse focusing or the specimen leaves the focus plane after focusing or the stage declines itself please...

Page 13: ...avoid damaging to both glasses and eyepieces 2 Open the eye cap if the user doesn t wear glasses so that it can prevent stray light disturbing the observation See Fig 20 3 7 Adjust the Objective Cente...

Page 14: ...g 23 Move the front lens of the condenser into light path when use the objective beyond 20X 3 Rotate the 20X objective into the light path and focus the specimen 4 Rotate the field diaphragm adjustmen...

Page 15: ...d of view then the field diaphragm is centered correctly See Fig 26 5 In actual use enlarge the field diaphragm a little to make its image circumscribed to the field of view For Transmitted Illuminati...

Page 16: ...the analyzer is shifted into the light path See Fig 29 2 Loosen the lock screw turn the analyzer turntable to change the vibrated direction of the analyzer A grid value of scale is 2 and the adjustin...

Page 17: ...an Bertrand lens and make the vibration direction of polarizer be vertical with the analyzer 3 14 Use the Compensator The compensator is inserted into the slot of the middle head in 45 declined The su...

Page 18: ...a then turn the coarse and fine focusing knobs until the image is clear 3 Rotate the Bertrand lens switch turntable locate it on position B to move the Bertrand lens into light path At that time the b...

Page 19: ...nermost for binocular observation pull the light path selecting pole to the outermost for trinocular observation TV and photography See Fig 35 3 19 2 Assemble and Use the TV Device 1 Loosen the lock s...

Page 20: ...pole to the outmost then operate the photography device according to its instructions 3 20 Use the Digital Viewing Head 3 20 1 Connect the Power Cord Insert the output connector of the transformer in...

Page 21: ...sabled when the white balance switch is popup while it is enabled when the switch is pressed down The switch is mainly for adjusting the color of the microscope TV image See Fig 40 When using the digi...

Page 22: ...k or not even Stain or dust has accumulated on the lens objectives or eyepieces Clean the lens Stains have accumulated on the specimen Clean the specimen 4 Stain or dust is observed in the view field...

Page 23: ...slide is not fixed correctly Adjust it correctly 5 Cannot move the slide smoothly The movable specimen holder is not fixed properly Fix it correctly 3 Electrical System No power supply Check the conn...

Page 24: ......