Chapter 5 Maintenance and Storage                                        A16.1065

 

5.1 Cleaning   

Clean  or  disinfect  the  lenses  and  other  components  according  to  the  following 
instructions

.

 

Cleaning tools 

 

Blower 

 

Soft brush 

 

Soft cotton cloth, lens cleaning tissue, gauze, etc. 

 

Absolute alcohol (ethyl alcohol or methyl alcohol), medical alcohol 

 

Petroleum benzene (only for wiping immersion oil) 

   

CAUTION

 

 

Petroleum  benzene  and  absolute  alcohol  used  for  cleaning  are  highly  flammable. 
Handle  them  with  due  care,  and  keep  them  away  from  fire  or  sparks  and  when 
turning the power switch on and off.   

 

When  handing  petroleum  benzene  or  absolute  alcohol,  always  follow  the 
instructions provided by the manufacturer.   

 

Do not use organic solvents (such as alcohol, ether, and thinner) when cleaning 
painted, plastic, or printed parts of this product. Using organic solvents might result 
in discoloration or cause printed text to fade. 

 

Use petroleum benzene only when wiping off immersion oil from objectives. Do not 
use petroleum benzene to wipe the prism surface of the eyepiece tube or filters. 

 

5.1.1 Cleaning the Lenses 

Keep the lenses free of dust, fingerprints, and other dirt. Any dirt on lenses and filters 
will  degrade  the  image.  If  the  lenses  become  dirty,  clean  them  according  to  the 
following procedure.   

 

Cleaning off minor dirt (such as dust)

 

(1)

 

Use a blower or similar tool to blow off dust. 

(2)

 

If the above method does not work, dust off using a soft brush, or gently wipe off 

with a gauze. 
   

 

Cleaning off heavy dirt (such as fingerprints or oil stains)

 

Wipe off dirt using a soft clean cotton cloth, lens cleaning tissue, or gauze moistened 
with a small amount of absolute alcohol. (ethyl alcohol or methyl alcohol). 
   

 

Tips for wiping 

Do not wipe the lens surface using the same portion of a cloth or tissue more than once.

 

 

5.1.2 Cleaning Parts other than Lenses

 

 

Cleaning off minor dirt (such as dust) 

Wipe off using a silicone cloth. 

Cleaning off heavy dirt (such as fingerprints or oil stains) 

Gently wipe off using a gauze moistened with a small amount of a diluted neutral 
detergent solution. 

Summary of Contents for A14.1065

Page 1: ...B G U Instruction Manual This instruction manual is intended for users of the Inverted Series Biological Microscope A16 1065 To ensure safety use perform the best performance and to familiarize yours...

Page 2: ...Specimen Placement and Stage Knob Adjustment 10 3 3 1 Mechanical Stage 10 3 3 2 Pallet 11 3 3 3 Auxiliary Stage 11 3 4 Adjusting the Interpupillary Distance 11 3 5 Focusing on the Specimens 12 3 5 1...

Page 3: ...5 Installing the Eyepiece 20 4 2 6 Installing the Camera 20 4 3 Assembly Episcopic Illumination Microscopy 21 4 3 1 Installing the Filter Cube 21 4 3 2 Installing the Fluorescence LED Unit 21 4 3 3 R...

Page 4: ...event lens glass to stick to fingerprint perspiration other factors influence observation prevent lens and other attachment drop damaged 2 Avoid placing the microscope in place with direct sunlight hi...

Page 5: ...e rear of the main body Do not hold any other parts such as the upper part of the illumination pillar focus knobs eyepiece tube or stage when carrying the microscope Doing so might result in dropping...

Page 6: ...use Cover up must wait for the lamp room to cool down 9 The microscope usage environmental requirements a For indoor use b The environment temperature range 10 35 c Maximum relative humidity 80 at 31...

Page 7: ...tive 4 Phase contrast objective 10 Phase contrast objective 20 Phase contrast objective 40 Infinity plane semi apochromatic objective 10 Infinity plane semi apochromatic objective 20 Infinity plane se...

Page 8: ...Magnification 4 10 20 40 10 20 40 numerical aperture N A 0 10 0 25 0 40 0 60 0 30 0 45 0 60 Work distance mm 30 10 2 12 2 2 7 4 8 3 6 Cover glass thickness 0 17 0 17 0 17 0 17 1 2 1 2 1 2 Conjugate d...

Page 9: ...A16 1065 1 2 Components The microscope A16 1065 components following diagrams A16 1065 Viewing head Camera mount LCD Mainframe Nosepiece Stage Condenser Objective Contrast slider Moving stage Eyepiec...

Page 10: ...nents following diagrams A16 1065 FL Viewing head Camera mount LCD Mainframe Fluorescence illumination Stage Condenser Filter Cube turret Fluorescent baffle Eyepiece Nosepiece Objective Moving stage U...

Page 11: ...adjustment knob 7 Condenser brightness 2 Condenser rotating knob 8 Fine focus knob 3 Condenser adjustment knob 9 Coarse focus knob 4 aperture diaphragm adjusting knob 10 Power indicator 5 Nosepiece 1...

Page 12: ...9 A16 1065 1 Bracket horizontal movement knob 2 Bracket vertical movement knob 3 Fluorescence cube turret A16 1063 3 1 2...

Page 13: ...careful when changing the objective The objective may collide with the specimen because of the short working distance of specimen observation 3 3 1 Movving stage The Movving stage is mounted and used...

Page 14: ...er 3 3 3 Auxiliary stage Connect the auxiliary stage to the left and right of the stage to expand the stage width After connecting the auxiliary stage the stage width is 300 mm 3 4Adjusting the Interp...

Page 15: ...or about 4 mm trip to limit The traveling distance of the objective for each knob is as follows Rotation of knob Distance traveled by objective Fine focus knob 1 rotation 0 2 mm Coarse focus knob 1 ro...

Page 16: ...result in an appropriate contrast and a favorable image The iris aperture adjusting handle is marked with the aperture of the diaphragm 3 7 Nosepiece Turn the nosepiece until it clicks in place stays...

Page 17: ...aphragm image with the phase ring image in the objective Decentering the phase ring and the PH annular diaphragm Basically the phase ring and the PH annular diaphragm should be adjusted so that they a...

Page 18: ...and switch the filter cube by rotating the filter cube switching unit 3 9 3 Filter cube switching unit Rotating the filter cube switching unit switches the filter cube When the filter cube is switched...

Page 19: ...on filter BP filter a barrier filter BA filter and a dichroic mirror DM Referring to the following items as a guide select a desired combination of filter cubes according to the characteristics of the...

Page 20: ...minute scale figures flash then click the button hour scale figures become flashing When the time figures flash three times it means setting successful Double click locking brightness or unlocking Whe...

Page 21: ...h component is shown in the following figure and the digital indicate the installation step Save the supplied hexagonal spanner You need it when you change parts CAUTION Before installation make sure...

Page 22: ...contrast slider horizontally into the condenser socket and notice the adjusting screw side of the phase contrast slider is facing the operator side Push into the process when you hear the click sound...

Page 23: ...itching the objective rotate the objective until it clicks to ensure that the desired objective enters the center of the optical path 4 2 5 Installing the Eyepieces Insert the eyepiece into the eyepie...

Page 24: ...scope and remove the cover 2 Mount the LED unit corresponding to the desired wavelength at the position that matches the address of the filter cube The back cover of the LED unit is marked with positi...

Page 25: ...ff immersion oil from objectives Do not use petroleum benzene to wipe the prism surface of the eyepiece tube or filters 5 1 1 Cleaning the Lenses Keep the lenses free of dust fingerprints and other di...

Page 26: ...disposal If a sample comes into contact with this product check the sample is hazardous or not If the sample is hazardous follow the standard procedure of your laboratory 5 2 Storage Store this produ...

Page 27: ...ced reversely Turn it over A lens the objective condenser eyepiece or collector is dirty Clean it The aperture diaphragm is too large Appropriate down Dirt or dust on the eyepiece Clean it The apertur...

Page 28: ...llary Distance does not adjust Adjust the Interpupillary Distance correctly the illumination is not appropriate Adjust illumination switch 6 3 Electrical System TROUBLE CAUSE SOLUTION The bulb cannot...

Reviews: