background image

 

21

                                                                                             

                     

       

5-2 

Assembling and operating

 

                                                                                       

5-2-1 

Assemble phase-contrast slider

 (See fig.29)

 

                                                                                            1.    Place  phase-contrast  slider

  upward  (on  letters 

side), insert it into the illumination holder from the 

right to the left.   

                                                                                            2.    Each  diaphragm  has  its  corresponding  part.  Enter 

the  diaphragm  into  optical  way  by  shifting  the 

phase-contrast until hearing a click.   

                                                                                            3.    Keep  the  adjusting  ring  of  aperture  diaphragm  in 

” when do the phase-contrast observation. 

 

             

                                                                       

5-2-2 

Center the light loop 

(See fig.30&31)

 

                                                                         

     

    XDSLP needs no centering 

                                                                                              1.    Place specimen on the stage, and focusing. 

                                                                                              2.    Remove  the  eyepiece,  and  replace  it  with  CT 

(centering telescope), insert in to the tube without 

diopter adjustment.   

                                                                                              3.    Make  sure  the  matched  phase  loop  (in  the 

phase-contrast objective) and the light loop (in the 

phase-contrast  slider)  is  shifting  into  the  optical 

way.   

                                                                                              4. While observe into the centering telescope, turn the 

fine  focusing  knob  which  corresponding  to  the 

light  loop

  to  focus  the  phase  loop

  of  the 

objective.(See fig. 30) 

                                                                                              5.    Insert  the  phase  center  adjusting  rod

  into  two 

holes

  on  the  phase-contrast  slider,  then 

adjusting  them  until  the  light  loop

  center 

overlay  with  phase  loop

  center.  (See  fig.30  & 

fig.31)

 

                                                                                              6.    Repeat the steps to center the other phase-contrast. 

Remember  to  use  the  same  phase-contrast  slider 

with  10X  and 20X.  Shift  the  optical  way  into  the 

objective  which  is  not  centered  to  see  if  the  light 

loop

  is away from the phase loop

. Repeat the 

centering steps if it is away. (See fig.30) 

 

 

 

 

Fig 29 

 

Fig 30 

 

Fig 31 

Summary of Contents for A14.0901-A

Page 1: ...To ensure the safety and obtain optimum performance and to familiarize yourself fully with use of this microscope we recommend that you study this manual thoroughly before operating the microscope Ret...

Page 2: ...g Device 10 4 1 Stand 10 4 2 Stage 11 4 3 Observing Tube 12 4 4 Illumination device 14 5 Phase contrast Observation 16 5 1 Nomenclature 16 5 2Assembling and operating 17 6 TV device Photography device...

Page 3: ...fuse It would be better to do so after cooling the bulb and the lamp holder Lamp selected only 6V 30W HAL bulb Philips 5761 7 It can adapt wide voltage from 90 to 240V There is no need for transformer...

Page 4: ...ty Sign Sign Signification It shows the surface gets hot and don t touch with bare hand Read the introduction before use Unsuitable operation would lead to person hurt or instrument faulty Main switch...

Page 5: ...LWDPH40 Long work distance condenser XDCD0 3 Phase contrast slider Phase center presetting slider XDSLP Phase center adjustable slider XDSL Circular central stage Metal stage XDCPM Glass stage XDCPG S...

Page 6: ...Stage fixed Stage extension board XDSGEX Attachable mechanical stage XDSGM Nosepiece fixed Quintuple nosepiece Illumination holder fixed Lamp holder XDAL2 Phase contrast slider Phase center presetting...

Page 7: ...part or touch the glass face Preserve the hexagon spanner and it will be used when changing the parts Halogen Bulb 6V 30W HAL Lamp holder XDAL2 Color Transmissive illumination group Color Eyepiece PL1...

Page 8: ...main power make sure the power switch is disconnect the power wire and wait until the lamp holder and bulb cool down 2 2 2 Assemble the transmission illumination groupware See fig 4 1 Unbolt the screw...

Page 9: ...us the sample by low magnification objective 4X or 10X when using the equipment Then change to the high magnification ones according as the situation observation requires Turn the condenser when chang...

Page 10: ...Insert the eyepiece into the tube thoroughly 2 2 8 Assemble the color filter Fig 11 Make sure the color filter is completely cool down before changing it Take down the filter holder and put in the co...

Page 11: ...to the supply socket and make sure the connection is well See fig 14 Use the special wire supplied by our company Choose another suitable wire if it s lost or damaged 90 240V wide voltage Connect the...

Page 12: ...ocation Phase contrast slide adopt phase center adjustable slide configured with phase center presetting rod Phase contract slider Color filter holder Diopter adjusting ring Interpupillary distance po...

Page 13: ...XD20 for their location Phase contract slider Color filter holder Phase center adjusting rod The adjusting ring of aperture diaphragm Interpupillary distance pointer Diopter adjusting ring Light path...

Page 14: ...nsion adjusting ring Make sure the coarse tension adjusting ring is being used to adjust the tension of coarse adjustment knob How to adjust tension Turn the coarse tension adjusting ring Coarse adjus...

Page 15: ...e when using 96 or 24 microtitration plate 2 In order to clip board of any type tightly please use holder and attachable mechanical stage as follows Use Terasaki holder which to place Terasaki board U...

Page 16: ...justment ring if the image is unclear when observing the left eyepiece with left eye See fig 21 The user could select any side of the eyepieces as a standard of reference if there has diopter adjustme...

Page 17: ...r observation tube Shift the optical way by moving the block of shifting the optical way Shift the block to the right until hearing a click when doing normal observation Shift the block to the left un...

Page 18: ...0 De infrared color filter use in photomicrograph to compensate the exposure time 4 4 2 Use the adjusting ring of aperture diaphragm See fig 25 The aperture diaphragm decides the numerical aperture of...

Page 19: ...26 Turn the parts below the condenser in clockwise to get larger operation space It could get the height of 150mm Remember it would not get well illumination when operating It only proper to use big P...

Page 20: ...ment 10X 20X light loop for 10X 20X Phase contrast objective 40X light loop for 40X Phase contrast objective hole for 45mm color filter See fig 27 5 1 3 XDSL phase contrast slider See fig 28 Phase cen...

Page 21: ...hout diopter adjustment 3 Make sure the matched phase loop in the phase contrast objective and the light loop in the phase contrast slider is shifting into the optical way 4 While observe into the cen...

Page 22: ...t bright image when seeing double light loop Light loop will deviate away from phase loop after moving or replacing a piece of thick specimen Repeat step 1 5 if it happens Repeat the centering steps i...

Page 23: ...on the trinocular head and get down the dust cover 2 Get down the dust cover caps of the TV adapter Insert adapter into the trinocular as the figure shows Screw on the lock screw tightly 3 Loosen the...

Page 24: ...he adapter to the joint of the assembly parts of camera Screw it tightly in clockwise See fig 36 5 Insert the photography eyepiece 3 2X to the photography tube then mount adapter and assembly parts of...

Page 25: ...by view finder when taking photo See instructions of the camera assembly parts for details 6 2 4 Adjust color temperature Use daylight type film to take photo 1 Place LBD color temperature transition...

Page 26: ...sion board 70 W 80 L mm attachable mechanical stage Work distance 120 X 80 Y mm Low coaxial focus adjustor attached with Petri dish holder Illumination Adopt 90 240V wide voltage the 6V 30W Halogen bu...

Page 27: ...ge phase contrast is out of work Unclear Image of fine structure Objectives is not place in the light path Turn the nosepiece into the right position until hearing a click 5 The aperture opened incorr...

Page 28: ...ot work No power supply Check the connection of the power cable 7 The bulb is not inserted correctly Insert it correctly 4 The bulb burnt out Replace it 4 2 The bulb burnt out usually Not use the spec...

Reviews: