3.Installation & Operation 

3.1 

Installation 

1) Put out the main body, loose the body fixed screw. 
2) Put out the reflection illumination unit and halogen lamp house, then fix the lamp house into the back of 
reflection illumination unit by the lamp house fixed screw. Insert the power wire of the lamp house into the 
power output socket of power house. Fix the reflection illumination unit to the main body, and tighten the 
body fixed screw. 
3) Put out diaphragm slide and polarizer slide. Put down a handle of diaphragm slide, and insert it into the 
groove of the back of the reflection illumination, put up the handle. Put on the polarizer slide as the same 
(position groove face to front and insert into reflection illumination) 
4) Put out binocular head (trinocular head), fix it onto the reflection illumination, and tighten the eyepiece 
tube fixed screw. 
5) Put out the dust cover of the eyepiece tube, and insert eyepiece into the tubes. 
6) Take out the objective from the packing box, and drive them into the holes of nosepiece orderly and 

tightly according to the times. 

7) If it is transmitting and reflection microscope, please put out condenser, and put it on the condenser 

holder. 

8) Put the glass plate stand on the stage, and fix it by the fixed screw, and place the glass plate to the stand. 
3.2 

Operation

 

Confirm the voltage is available, then take power 
1)

 

Turn potentiometer, and make the light to the lowest position. Open the power switch, and adjust the 
potentiometer to get the available light. If reflection observing, operate the potentiometer and power 
switch in the power house; if transmitting observing, operate the potentiometer and power switch in the 
microscope main body. 

2)

 

Sample placing     
Choose an available stage plate, put the sample onto it, turn the stage knob to put the sample into the 
light path. If reflection observing, stage plate and glass plate should be chosen; if transmitting 
observing, glass plate and clamp should be chosen. 

3)    Put 10X objective into light path, turn coarse focusing knob, observe from right eyepiece with right 

eye, then turn slowly fine focusing knob to make the image clear after finding image. 

4)    Interpupillary distance adjustment 

Interpupillary distance is different for everyone, so interpupillary distance should be adjusted before 
using binocular microscope. Please take the two eyepiece tubes to turn until the bright rings observed 
by two eyes are in superposition entirely. 

5)    Diopter adjustment   

 

As focusing for binocular, user should observe right eyepiece with right eye, and make the right eyepiece 

clear by focusing adjustment, then observe the left eyepiece, at the same time, adjust the diopter ring of the 
left eyepiece tube to make the image of left eyepiece clear as same as the right eyepiece. 
6) Field diaphragm adjustment 

a

. If reflection observing, the field diaphragm has been adjusted well in factory, and user doesn’t need to 

adjust. Using field diaphragm lever to adjust it. 

b

. If transmitting observing, turn the field diaphragm ring to make the diaphragm smaller than the field of 

eyepiece, the image of field diaphragm can be gotten in field. Adjust the centering adjust screw to make the 
center of field diaphragm overlap with the center of eyepiece field. Make the image of field diaphragm 
smaller than eyepiece field to adjust again. Larger the field diaphragm to make it disappeared.   

Summary of Contents for A13.2604-A

Page 1: ...604 B Metallurgical Microscope Reflect Transmit Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument ...

Page 2: ...in use Alcohol gasoline paper and other flammable materials can t near the lamp in case of fire 5 Notes On Replacing The Bulb Replacement should be based on the identity of the instrument using the same specifications of the bulb otherwise it may cause equipment damage The power supply must be cut off before bulb replacement the bulb must be cooled off completely before proceeding 6 Carry Power mu...

Page 3: ...57X C mount also provide digital camera for CCD SLR digital camera or special designing camera by our company It has many advantages for example fine appearance stable performance excellent image convenient usage etc Widely used in factories universities and relative research institutes This series microscope is suitable for observing various metallurgical specimens casting quality and identificat...

Page 4: ...topf trinocular head with output connector for CCD Camera SLR Digital Camera 1 6 Nosepiece Quintuple 1 7 Mechanical stage For Reflecting Illumination Size 200mm 150mm Travel range 77mm 52mm For Transmitting Illumination Size175mm 145mm Travel range 77mm 52mm Two slid holder 1 8 Focusing systems Coaxial Coarse and Fine Focusing Knobs Coarse stroke 22mm Fine division 2µm Condenser up down range 25mm...

Page 5: ...Plate Stage Halogen lamp house Lamp house fix screw Reflection illumination Aperture diaphragm lever Filter slider Field diaphragm lever Polarizing slider Main body Right fine focusing knob Right coarse focusing knob Ground glass slider Slider handle Eyepiece tube fix screw Power line with plug Halogen Lamp Centering Knob MIT300 Upright Metallurgical Microscope ...

Page 6: ... switch Fuse Input socket Output socket MIT300 Upright Metallurgical Microscope Camera adaptor with dust cover for trinouclar head tube Handle Stage up down adjusting screw Stage plate fix screw Coarse tension adjusting knob Left coarse focusing knob Left fine focusing knob ...

Page 7: ...izer adjusting ring and fixed screw Field diaphragm adjusting ring Condenser fixed screw Potentiometer MIT500 Transmitting Reflecting Metallurgical Microscope Clamp locking screw Power switch Power socket and Fuse Adjustable Level for swing out Condenser Condenser focusing knob ...

Page 8: ...wer house if transmitting observing operate the potentiometer and power switch in the microscope main body 2 Sample placing Choose an available stage plate put the sample onto it turn the stage knob to put the sample into the light path If reflection observing stage plate and glass plate should be chosen if transmitting observing glass plate and clamp should be chosen 3 Put 10X objective into ligh...

Page 9: ...wn adjusting screw after get the available position 10 Usage of 100X immersion oil objective Adding moderate immersion oil between the front lens of 100X objective and the cover glass of specimen can make the image clearer Please pay attention that air bubble and impurity can t be in the immersion oil otherwise the image would be affected First of all take 40X objective which has been focused well...

Page 10: ...avoid fungus and mist damage instrument 4 2 4 Please pay attention to prevent microscope from violent shake and vibration in application and in carrying Don t drag it on the surface of worktable to avoid damage to microscope and worktable 4 3 Replacement of bulb 4 3 1 Turn off power and pull out plug 4 3 2 Wait the bulb become cool Please be sure that the bulb is cool then follow by the nest opera...

Page 11: ...lass 0 17mm Cover glass of specimen isn t in up direction Place specimen correctly Surface of objective lens Is dirty especially it is easy for the front lens of 40X objective to dip in immersion oil Clean it Immersion oil isn t used for 100X objective oil Use immersion oil Immersion oil doesn t meet the requirement Use immersion oil supplied by us There is bubble in immersion oil Clear the bubble...

Reviews: