36
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1 – Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11; Group 1, Class B
Home healthcare environment
Harmonic distortion
IEC 61000-3-2; Class A
Home healthcare environment
Voltage fluctuations and flicker IEC 61000-3-3; Compliance
Home healthcare environment
Table 2
Enclosure Port
Phenomenon
Basic EMC standard
Immunity test levels
Home Healthcare Environment
Electrostatic Discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM field
IEC 61000-4-3
10 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
Refer to table 3
Rated power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Table 3
Proximity fields from RF wireless communications equipment
Test frequency (MHz)
Band (MHz)
Immunity test levels
Professional healthcare facility environment
385
380-390
Pulse modulation 18 Hz, 27 V/m
450
430-470
FM, ±5 kHz deviation, 1 kHz sine, 28V/m
710
704-787
Pulse modulation 217 Hz, 9 V/m
745
780
810
800-960
Pulse modulation 18 Hz, 28 V/m
870
930
1720
1700-1990
Pulse modulation 217 Hz, 28 V/m
1845
1970
2450
2400-2570
Pulse modulation 217 Hz, 28 V/m
5240
5100-5800
Pulse modulation 217 Hz, 9 V/m
5500
5785