Optima Scale OP-918 User Manual Download Page 15

13

Function

Parameter Settings/Options

Negative Display

C25

0 = -9d
10 = -10% max. capacity
20 = -20% max. capacity
50 = -50% max. capacity
100 = -100% max. capacity

Standstill Time

C26

0 =  quick

1 =  medium

2 =  slow

Standstill Range

C27

1 = 1d                                                         d = division

2 = 2d 
5 = 5d 
10 = 10d

Digital Filter

 

(for filtering moving 

weight such as  
animals)

C28

0 = turn off dynamic filter
1 = 1 digital filter strength
2 = 2 digital filter strength
3 = 3 digital filter strength
4 = 4 digital filter strength
5 = 5 digital filter strength
6 = 6 digital filter strength
Note: The higher the number, the higher the filter strength

Noise Filter

C29

0 = turn off noise filter
1 = 1 digital filter strength
2 = 2 digital filter strength
3 = 3 digital filter strength

Gravity of Calibration 
Location

C36

9.7000 - 9.9999

Gravity of Destination

C37

9.7000 - 9.9999

Version No.

C38

Print Mode

C41

0 = auto mode

1 = gross mode

2 = tare mode

Print Carriage Return

C42

0 - 9 (How much space between print outs)

Space Print

C43

0 - 9  (Where the data prints on the paper: 0 = left ; 9 = right)

Date Print

C44

0 = do not print the date

1 = print the date

Time Print

C45

0 = do not print the time

1 = print the time

Summary of Contents for OP-918

Page 1: ...optimascale com OP 918 INDICATOR USER S MANUAL Pallet Jack Indicator...

Page 2: ...rations and Set Up 1 Features 2 Specifications 3 Power Supply 4 Display and Key Descriptions 5 Operating Instructions 6 8 Calibration 9 10 Indicator Parameter Settings 11 14 Helpful Definitions 15 Con...

Page 3: ...e sure the weight is not over the Max capacity as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the s...

Page 4: ...ature Accumulation weighing Printer optional Splash proof keyboard and display Power saving mode NTEP approved for 5 000 divisions Rechargeable battery Technical Parameters Stimilating voltage 3 3VDC...

Page 5: ...3 SPECIFICATIONS OP 918 OP 918M...

Page 6: ...charged 7 Indicator charge lasts up to 22 hours Battery Please charge the internal battery fully before first time use A 110 to 220V AC adapter should be provided with your indicator Please use only...

Page 7: ...OFF to enter into calibration 3 Combined with the TOTAL key to check the Accumulative sum lb The weight is shown in pounds kg The weight is shown in kilograms hold Shows you are in Hold mode Gross Sho...

Page 8: ...to tare out the weight Tare Function The Tare function is used when you only wish to see the current change in weight not the entire amount of weight that is on the scale When the indicator is in gros...

Page 9: ...adding together will flash on the display followed by the total The total will display by flashing between 2 sets of numbers There are 8 digits in total the display will flash 4 at a time the first 4...

Page 10: ...to Hold If the weight on the scale is above 20d 20 x division and is stable the indicator will hold that weight on the screen for 3 seconds then go back to general weighing Pressing the hold key is un...

Page 11: ...this scale by using TOTAL to move the cursor to the right and UNIT and ZERO to move the values down and up The C4 channel is used to enter in the max capacity of the scale Make sure to enter the corre...

Page 12: ...the calibration weight on the scale 29 If the scale does not show the value of the calibration weight check that the pallet jack is level on the ground and on a flat surface and recalibrate 30 Unload...

Page 13: ...nit C01 1 kg 2 lb 3 gram 4 oz Note for calibration only kg or lb are allowed Decimal Setting C02 0 no decimal 1 0 0 2 0 00 3 0 000 4 0 0000 Division Graduation Setting readability of the least signifi...

Page 14: ...4 Average hold for animal weighing averages the weight from a sample of 3 seconds Inner Code Display C15 check the inner code raw data Set Date C16 Set date from left to right year month day Set Time...

Page 15: ...5 5 digital filter strength 6 6 digital filter strength Note The higher the number the higher the filter strength Noise Filter C29 0 turn off noise filter 1 1 digital filter strength 2 2 digital filte...

Page 16: ...Alarm C14 000000 Inner Code Display C15 Set Date C16 Set Time C17 Communication Setting C18 0 Baud Rate C19 3 9600 Manual Zero Range C20 10 Initial Zero Range C21 10 Zero Tracking C22 0 5 Zero Trackin...

Page 17: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...

Page 18: ...s of 350 at most 4 wire or 6 wire load cell connections are both okay Please contact us directly if you have other special needs for your application There are two connection methods between the load...

Page 19: ...See Q A section ERR1 During calibration weight is not used or the weight is above the max capacity Use correct weight within the defined range ERR2 During calibration the weight is below the minimum r...

Reviews: