Optima Scale OP-918 User Manual Download Page 11

9

1.  Turn on the scale by holding ON/OFF

 

 for 2 seconds. 

2.  Press and hold SET

 

 

and ON/OFF

   

together to access the setup menu.

3.  If done correctly, the display should now show 

C01

.

4.  Press SET

 

 to access the C1 channel. The display should show [

C1 

 

1

].

5.  Press ZERO   to choose which unit you want to calibrate in (1 = kg, 2 = lb).
6.  Press SET

 

 to set the value. The display will now show 

C02

7.  Press SET

 

 to access the C2 channel. The display should show [

C2 

 

0

].

8.  Press ZERO 

 

to change the setting to the decimal places desired (The C2 channel is 

used to adjust the decimal point on the scale. A value of 1 means there is one digit 
behind the decimal point.)

9.  Press SET

 

 to set the value. The display will now show 

C03

.

10. Press SET

 

 to access the C3 channel. The display should show [

C3

   

5

].

11. Press ZERO 

 

to cycle through the values until the desired graduation appears.  

(The C3 channel adjusts the division of the scale. A value of 1 selected and C2 set to 
1, the scale will read in 0.1 lb. increments.) 

12. Press SET

 

 to set the value. The display will now show 

C04

.

13. Press SET

 

 to access the C4 channel. The display will show [

021000

].

14. Enter in the maximum capacity you want to use for this scale by using TOTAL   to 

move the cursor to the right, and UNIT   and ZERO 

 

to move the values down and 

up. (The C4 channel is used to enter in the max capacity of the scale; Make sure to 
enter the correct max capacity of your scale, most pallet jack’s are 5,000 lbs.) 

15. Press SET

 

 to set the value. The display will now show 

C05

.

16. Press SET

 

 to access the C5 channel. The display should show [

C5  0

].

17. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
18. Press ZERO 

 

to change the value to 1.

19. Press SET

. The display will count down from 10-1 while the scale is  

calibrating zero. When the display shows 0 the zero calibration is complete.

20. Press SET

 

 to continue. The display will now show 

C06

.

21. Press SET

 

 to access the C06 channel. The display will show [

C6  0

].

22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO   to 

set the value of C6 to [

C6  1

]. Press SET

 

. The display will flash 

SPAN

, and 

then show [

010170

].

23. Enter the value of the calibration weight you will use (at least 10% of max capacity 

you set in C04) by using TOTAL   to move the cursor right, and UNIT    and ZERO   
move the values down and up.

24. Place the calibration weight you have on the empty scale and press SET

.

25. The scale will count down from 10 to 0. Once 0 has been reached, the display will 

show 

CALEnd

.

OP-918 PALLET JACK

CALIBRATION PROCEDURE

Summary of Contents for OP-918

Page 1: ...optimascale com OP 918 INDICATOR USER S MANUAL Pallet Jack Indicator...

Page 2: ...rations and Set Up 1 Features 2 Specifications 3 Power Supply 4 Display and Key Descriptions 5 Operating Instructions 6 8 Calibration 9 10 Indicator Parameter Settings 11 14 Helpful Definitions 15 Con...

Page 3: ...e sure the weight is not over the Max capacity as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the s...

Page 4: ...ature Accumulation weighing Printer optional Splash proof keyboard and display Power saving mode NTEP approved for 5 000 divisions Rechargeable battery Technical Parameters Stimilating voltage 3 3VDC...

Page 5: ...3 SPECIFICATIONS OP 918 OP 918M...

Page 6: ...charged 7 Indicator charge lasts up to 22 hours Battery Please charge the internal battery fully before first time use A 110 to 220V AC adapter should be provided with your indicator Please use only...

Page 7: ...OFF to enter into calibration 3 Combined with the TOTAL key to check the Accumulative sum lb The weight is shown in pounds kg The weight is shown in kilograms hold Shows you are in Hold mode Gross Sho...

Page 8: ...to tare out the weight Tare Function The Tare function is used when you only wish to see the current change in weight not the entire amount of weight that is on the scale When the indicator is in gros...

Page 9: ...adding together will flash on the display followed by the total The total will display by flashing between 2 sets of numbers There are 8 digits in total the display will flash 4 at a time the first 4...

Page 10: ...to Hold If the weight on the scale is above 20d 20 x division and is stable the indicator will hold that weight on the screen for 3 seconds then go back to general weighing Pressing the hold key is un...

Page 11: ...this scale by using TOTAL to move the cursor to the right and UNIT and ZERO to move the values down and up The C4 channel is used to enter in the max capacity of the scale Make sure to enter the corre...

Page 12: ...the calibration weight on the scale 29 If the scale does not show the value of the calibration weight check that the pallet jack is level on the ground and on a flat surface and recalibrate 30 Unload...

Page 13: ...nit C01 1 kg 2 lb 3 gram 4 oz Note for calibration only kg or lb are allowed Decimal Setting C02 0 no decimal 1 0 0 2 0 00 3 0 000 4 0 0000 Division Graduation Setting readability of the least signifi...

Page 14: ...4 Average hold for animal weighing averages the weight from a sample of 3 seconds Inner Code Display C15 check the inner code raw data Set Date C16 Set date from left to right year month day Set Time...

Page 15: ...5 5 digital filter strength 6 6 digital filter strength Note The higher the number the higher the filter strength Noise Filter C29 0 turn off noise filter 1 1 digital filter strength 2 2 digital filte...

Page 16: ...Alarm C14 000000 Inner Code Display C15 Set Date C16 Set Time C17 Communication Setting C18 0 Baud Rate C19 3 9600 Manual Zero Range C20 10 Initial Zero Range C21 10 Zero Tracking C22 0 5 Zero Trackin...

Page 17: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...

Page 18: ...s of 350 at most 4 wire or 6 wire load cell connections are both okay Please contact us directly if you have other special needs for your application There are two connection methods between the load...

Page 19: ...See Q A section ERR1 During calibration weight is not used or the weight is above the max capacity Use correct weight within the defined range ERR2 During calibration the weight is below the minimum r...

Reviews: