6.3  Care of Fiber-optic Connectors 

 

Damage  to  optical  connectors  account  for  more  than  70  percent  of 
equipment  performance  degradation.  To  avoid  such  damage,  the  user 
should  use  only  industrial  grade  99%  pure  isopropyl  alcohol  and  follow 
the procedures below to keep the connectors, adaptors and receptacles 
clean. 

 

Cleaning Optical Connector End-face with Wipe and Alcohol 
 

To  properly  clean  optical  connectors  utilizing  lens  tissue  grade  wipes  and  alcohol 
follow  the  procedure  below.  The  moist  wipe  removes  dust  particles,  oil  and 
contaminants  that  may  damage  or  blot  the  end-face  of  the  connector  during 
connection.  The  dry  wipe  removes  residual  alcohol  that  may  be  ignited  by  optical 
emission. 
 

1.  Disable the optical output and turn off unit to prevent accidental exposure or 

damage to the optical connector by optical emission. 

2.  Moisten  a  wipe  with  alcohol  by  placing  on  top  of  the  alcohol  dispenser  and 

push down to saturate the wipe. 

3.  Place the moist wipe on a work surface, and place a second dry wipe next to 

it. 

4.  Wipe  the  optical  connector,  end-face  down  on  the  moist  wipe  3  times  and 

then repeat on the dry wipe. 

5.  Visually  inspect  the  end-face  of  the  optical  connector  with  an  optical 

microscope to verify cleanliness. Repeat steps 2 to 5 as needed. 

 

Cleaning Optical Connector Sides, Receptacles, Adaptors with Swab and 
Alcohol 

 
Dust or particles can adhere to the insides of receptacles and adaptors or the sides 
of the optical connector ferrule. Their presence can affect the alignment of the optical 
fiber connectors and increase connection loss. To properly clean optical connectors, 
receptacles, and adaptors utilizing a swab and alcohol follow the procedure below: 
 

1.  Disable the optical output and turn off unit to prevent accidental exposure or 

damage to the optical connector by optical emission. 

2.  Moisten the swab by placing it on top of the alcohol dispenser and push down 

to saturate the swab. 

3.  For  receptacles,  adapters,  or  other  connection  points,  insert  the  moistened 

swab  and  rotate  the  tip  1/2  turn  clockwise  and  counter-clockwise  6  times 
while applying light but firm pressure.  

4.  For  fiber  connectors,  rotate  the  tip  of  the  moistened  swab  5  revolutions 

around the connector while applying light but firm pressure. 

5.  Visually  inspect  the  end  face of  the  connector  with an  optical microscope  to 

verify cleanliness. Clean end-face as needed. 

 

Summary of Contents for TWL-L-R Series

Page 1: ... must read this manual before operating the TWL L R unit Operations other than those described in this manual may result in personal injury and or damage to the unit Note that any attempt to open or fix the equipment without prior approval by Optilab LLC voids the warranty Ver 2 1 November 21 2015 ...

Page 2: ...nd control revision 2 1 11 21 15 Software operation revision Copyright 2015 by Optilab LLC All rights reserved This document is copyrighted property of Optilab LLC It may not be used in whole or in part for manufacture sale or design of items without the written permission of Optilab LLC Information herein is preliminary and subject to change without any notices ...

Page 3: ..._________________________________________________ 4 2 5 Software Interface and Control ______________________________________________ 4 2 5 1 Serial communication ___________________________________________________________4 2 5 2 Operating Guide for GUI _________________________________________________________5 2 5 3 Installation ____________________________________________________________________...

Page 4: ...w laser line width high Side Mode Suppression Ratio SMSR excellent wavelength stability and optical output level exceeding 20mW With the standard Polarization Maintaining PM fiber output TWL L R offers a linearly polarized output These features make TWL L R well suited for many different applications including DWDM device testing optical sensing and laboratory measurement applications 1 3 Features...

Page 5: ... result in damage to the surface and internal circuits 2 Operation 2 1 Introduction This chapter describes how to operate the TWL L R unit and discusses the location and function of the controls and connectors 2 2 Initial Inspection Your TWL L R unit was carefully inspected before it left the manufacturer It should be in proper working order upon receipt You should however inspect the unit for any...

Page 6: ...e green when the device has received proper AC power connections and the rear panel AC switch is enabled The individual LED indicators beneath each optical output port will illuminate when an active output is detected Laser Enable Switch This key switch toggles the internal TWL L R laser source On and Off 5 LCD display This LCD display shows the power level and wavelength of the internal L Band tu...

Page 7: ...ection for more adjustment and operation details Patch Cord Swapping Procedure 1 Toggle the front panel Laser Enable Switch to the Off position to disable the TWL L R module output 2 Swap patchcords as desired Only connect the indicated connector patchcords to the optical input output receptacles cleaning them as necessary 3 Toggle the front panel Laser Enable Switch to the On position normal oper...

Page 8: ... DVD which is supplied by Optilab together with the TWL L R package 2 5 3 Installation Explore the software DVD included and utilize the RS232 to USB drivers for your system to recognize the attached TWL L R device And use the TWL Software Package to TWL L SETUP file to setup the LabView GUI for operation If LabVIEW is not already installed on your PC you will need to use the LabVIEW 2012 RTE 32bi...

Page 9: ...g operation 3 Optical Settings Adjustment Use the box on the left side of the settings section of the screen to select the optical adjustment Note that only the selected box on the left will be adjusted at any given time Set PWR Optical power of the system from 6 16 dBm Set Wavelength Center wavelength in 0 001nm accuracy please refer to test data sheet for set range Set Switch Turns the output la...

Page 10: ... setting is properly communicated 5 To close the software select the stop button this should automatically disable the LabVIEW program from operation 2 6 LCD Front Panel Controls The Optilab TWL L R device allows user to adjust the lasers by using 5 buttons on the front panel directly as shown below next to the display Once the TWL L R device has been powered on and the key switch has been turned ...

Page 11: ... Optilab LLC 600 E Camelback Road Phoenix AZ 85012 Phone 602 343 1496 Fax 602 343 1489 Email sales oequest com enable disable the tunable LD or use the Left Right or Middle button to edit the power or wavelength will be both functional ...

Page 12: ...sure optical output power with power meter and compare with original test data Return to Optilab for repair if the difference is high 4 dB and cannot be corrected by cleaning or replacing the optical connectors Always apply dust cover plugs to unused optical receptacles to prevent the damage of internal optical connectors Optical output power unstable C Insufficient optical output isolation S Conn...

Page 13: ...z w SBS Carrier to Noise Ratio CNR 50 dB typ 5 dBm Side Mode Suppression Ratio 55 dB typ Relative Intensity Noise RIN 145dB Hz Polarization Extinction Ratio 20 dB min Optical Isolation 30 dB min Fiber Type Panda 1550 PM Fiber 5 Mechanical Specifications Power Supply Requirements 100 240 VAC Optical Connectors PM Narrow Key FC APC Standard additional types available Operating Temperature 0 C to 40 ...

Page 14: ...al or consequential loss Note that the warranty will be void upon any attempt to open or to fix the equipment by the user without prior approval of Optilab LLC 6 2 Service and Calibration Your TWL L R unit has been designed to provide years of trouble free operation No internal maintenance is required provided that the equipment is properly handled operated and kept away from contamination For any...

Page 15: ...pe next to it 4 Wipe the optical connector end face down on the moist wipe 3 times and then repeat on the dry wipe 5 Visually inspect the end face of the optical connector with an optical microscope to verify cleanliness Repeat steps 2 to 5 as needed Cleaning Optical Connector Sides Receptacles Adaptors with Swab and Alcohol Dust or particles can adhere to the insides of receptacles and adaptors o...

Reviews: