![OPTILAB NLFS-L-B User Manual Download Page 9](http://html1.mh-extra.com/html/optilab/nlfs-l-b/nlfs-l-b_user-manual_1642557009.webp)
NLFS-L-
B User’s Manual
Page 6 of 13
Optilab, LLC
600 E. Camelback Road, Phoenix, AZ 85012
Phone: 1-602-343-1496, Fax: (602) 343-1489, Email: [email protected]
4.
Optical Power Setting Page
This page offers the function to set the laser power. It also displays the
minimum and maximum optical power the laser can output. The optical
power setting resolution is 0.01 dBm. Use the SHIFT button to select the
digit of interest, use the INC to change the value of the digit, and then use
the SAVE button to save the setting. Wait 5 sec for the setting to be
updated. Use the BACK button to return to the Laser Information Page.
5.
Low Noise Mode Setting Page
This page offers the function to set noise mode. Press SHIFT to activate
the setting, then use INC button to set the low noise mode on or off, use
the SAVE button to save the setting. Wait 5 sec for the setting to be
updated.
Note
: The laser outputs narrow linewidth only when Low Noise Mode is ON.
Note
: Setting the Low Noise Mode to OFF each time when the laser is
powered on, then setting the Low Noise Mode to ON or OFF depends on the
requirement of the use.