![OPTILAB NLFS-L-B User Manual Download Page 13](http://html1.mh-extra.com/html/optilab/nlfs-l-b/nlfs-l-b_user-manual_1642557013.webp)
NLFS-L-
B User’s Manual
Page 10 of 13
Optilab, LLC
600 E. Camelback Road, Phoenix, AZ 85012
Phone: 1-602-343-1496, Fax: (602) 343-1489, Email: [email protected]
3. Place the moist wipe on a work surface and place a second dry wipe next to
it.
4. Wipe the optical connector, end-face down on the moist wipe 3 times and
then repeat on the dry wipe.
5. Visually inspect the end-face of the optical connector with an optical
microscope to verify cleanliness. Repeat steps 2 to 5 as needed.
Cleaning Optical Connector Sides, Receptacles, Adaptors with Swab and
Alcohol
Dust or particles can adhere to the insides of receptacles and adaptors or the sides
of the optical connector ferrule. Their presence can affect the alignment of the optical
fiber connectors and increase connection loss. To properly clean optical connectors,
receptacles, and adaptors utilizing a swab and alcohol follow the procedure below:
1. Disable the optical output and turn off unit to prevent accidental exposure or
damage to the optical connector by optical emission.
2. Moisten the swab by placing it on top of the alcohol dispenser and push down
to saturate the swab.
3. For receptacles, adapters, or other connection points, insert the moistened
swab and rotate the tip 1/2 turn clockwise and counterclockwise 6 times while
applying light but firm pressure.
4. For fiber connectors, rotate the tip of the moistened swab 5 revolutions
around the connector while applying light but firm pressure.
5. Visually inspect the end face of the connector with an optical microscope to
verify cleanliness. Clean end-face as needed.