SYMPTOMS
CORRECTIVE ACTION
Probe freezing
The probe does not freeze properly
a)
Incorrect operating pressure. Replace the cylinder.
b)
Humidity in the probe. Wait for the end of the
cleaning cycle (Purge).
c)
Dirty probe filter. Send for repair.
d)
The filters of the console are dirty. Contact the
Optikon technical service.
e)
The cylinder valve is closed. Open the valve of the
cylinder.
f)
The cylinder is empty. Replace the cylinder.
g)
A siphon cylinder is installed. Replace with a
cylinder without siphon.
Probe defrosting
The probe does not defrost
a)
Defective valve. Contact the Optikon technical
service.
Errors
From Err 01 to Err 05
Contact the Optikon technical service or authorised
personnel.
Err 06
High pressure. Do not use the instrument in this
condition. Check the type of gas used or the operating
pressure of the system to which the OPTIKON Cryo-
Line is connected.
Err 07
Low pressure. Do not use the instrument in this
condition. Replace the cylinder or check the operating
pressure of the system to which the instrument is
connected.
Err 08
The footswitch has been pressed without the probe
connected. Connect the probe before pressing the
footswitch.
Err 09
Accidental probe disconnection. Contact the Optikon
service centre or authorised personnel.
Err 10
Thermocouple disconnection or failure. Check that
the thermocouple connector is connected to the
instrument. If problems continue, contact the Optikon
service centre or authorised personnel.
Err 11
Preset temperature cannot be reached. Check the
pressure of the pneumatic circuit. If the pressure is
less than 40 bar, replace the cylinder. Perform a
probe cleaning cycle.