CRYO-LINE
Manual for installation and use
OPTIKON 2000
Code 131002 EN
2014-02-06 Rev.F
8-5
8.5
SYSTEM TEST
User Interface
Operation
At the start-up of the instrument it is possible to carry out a test cycle to assess the
efficiency and seal of the pneumatic circuit electrovalves.
If you wish to avoid the initial test, press the key next to the writing SKIP on the display.
In any case, it is advisable to always carry out the initial test, following this procedure:
1.
Connect the cylinder to the instrument, as explained in the installation
procedure.
2.
Switch on the instrument.
3.
A sound will be heard a few seconds after the start-up, as an assessment test of
the audio circuit.
4.
If the gas pressure at the inlet is too low or too high, a message will appear on
the display. Adjust the reducer located on the cylinder or, in the case of pressure
that is too low, replace the cylinder. The attainment of the ideal pressure will be
accompanied by a sound signal and the disappearance of the error message.
5.
Insert a cryosurgical probe into the probe connection.
6.
If operations 1-5 have been carried out properly, start the circuit test.
7.
If one of these tests fails, contact the service centre or your Optikon distributor.