background image

Page 17

One side of the image is out of focus.

The nosepiece is not in the 

center of the light path

Turn the nosepiece to a click stop

The specimen is out of place 

(tilted)

Place the specimen flat on the stage.

The optical performance of the 

sample cover glass is poor

Use a cover glass of better quality

II. Mechanical Section:

The coarse focus knob is hard to 

turn.

The tension adjustment collar is 

too tight

Loosen the tension adjustment collar

The focus is unstable.

The tension adjustment collar is 

too loose

Tighten the tension adjustment collar

III. Electric section

The LED doesn’t turn on.

No power supply

Check the power cord connection

The brightness is not enough

The brightness adjustment is 

low

Adjust the brightness

The light blinks

The power cord is poorly con

-

nected

Check the power cord

IV. Viewing tube assembly

The field of view of the two eyes is 

different

The interpupillar distance is not 

correct

Adjust the interpupillar distance

The dioptric correction is not 

right

Adjust the dioptric correction

The viewing technique is not 

correct, and the operator is 

straining the eyesight

When look into the objective, do not stare 

at the specimen but look at the whole field 

of view. Periodically, move the eyes away 

to look at a distant object, then back into 

the objective

V. Microphotography and video

The image is unfocused

Incorrect focussing

Adjusting the focus system as in the pres

-

ent manual

The edge of the image is unfocussed

To some degree, it is inherent 

to the nature of achromatic 

objectives

The problem can be minimized by a correct 

setting of the aperture diaphragm

Bright patches appear on the image

Stray light is entering the 

microscope through the eye

-

pieces and through the camera 

viewfinder

Cover the eyepieces and the viewfinder 

with a dark cloth

Summary of Contents for IM-2

Page 1: ...Version 1 Issued 30 05 2016 Model IM 2 IM 2ERGO IM Series INSTRUCTION MANUAL...

Page 2: ...e of Contents Warning Symbols and conventions Safety Information Intended use List of accessories and spare parts Overview Unpacking Assembling Using the microscope Maintenance Troubleshooting Equipme...

Page 3: ...s manual CAUTION This symbol indicates a potential risk and alerts you to proceed with caution ELECTRICAL SHOCK This symbol indicates a risk of electrical shock Safety Information Avoiding Electrical...

Page 4: ...romatic for phase contrast 40x 0 60 to use with M 776 M 776 Phase ring 40x to use with M 774 M 778 C mount adapter for 1 3 1 2 and 2 3 sensor M 036 Dust cover type 7 M 795 Fluorescence attachment HBO1...

Page 5: ...TRAST SLIDER TRINOCULAR VIEWING TUBE FIXED DIOPTRIC ADJUSTMENT RING EYEPIECE LED HOUSING CONDENSER COARSE FOCUS KNOB FINE FOCUS KNOB TENSION ADJUSTMENT COLLAR OBJECTIVE STAGE MICROSCOPE BODY LIGHT ADJ...

Page 6: ...and eyepieces fall out and get damaged Using both hands one around the arm and one around the base lift the microscope from the container and put it on a stable desk Assembling Once you open the box t...

Page 7: ...gh the stage opening Fig 4 Clean the objectives regularly In inverted microscopes the objectives are very sen sitive to dust To prevent dust and contamination from entering the microscope cover all th...

Page 8: ...ide of the stage to enlarge the working sur face The mechanical stage must be installed on the side opposite the extension For right handed operators the mechanical stage is normally installed on the...

Page 9: ...as the whole thickness is less than 11 mm Installing the color filters Fig 13 Fig 14 Fig 10 Fig 11 Fig 12 1 Insert the condenser illumination unit into the bracket Fig 10 2 Turn the condenser illumina...

Page 10: ...lost or damaged please refer to qualified service Connect the power cord to a grounded earthed power supply only Before replacing the fuse turn the main switch to O off and unplug the power cord Rota...

Page 11: ...aki and Petri diameter 65mm M 793 3 Holder for slide and Petri diameter 54mm M 793 4 Holder for 2 2 slides M 793 6 Holder for Uterm hl Chamber M 793 3 needed M 793 8 Load bearing side extension for IM...

Page 12: ...them around their common axis until the fields of view co incide The graduation on the interpupillary distance indicator pointed by the spot on the eyepiece holder shows the distance between the opera...

Page 13: ...ure diaphragm lever clockwise Fig 23 ILLUMINATION UNIT Phase contrast slider Adjustable phase slider The light ring is pre centered when the microscope leaves the factory It should therefore need no f...

Page 14: ...the camera to the ring Warning for some cameras mainly re flex the ring is not included with the mi croscope and it should be supplied by the user For the photography of dark specimens obscure the eye...

Page 15: ...he equipment and that the lamp switch is in off position Users should observe all safety regulations of the region The equipment has acquired the CE safety label However users do have full responsibil...

Page 16: ...ragm is too small Open the aperture diaphragm Poor image quality The image is not sharp The contrast is not high The details are not clear The phase contrast is low The nosepiece is not in the center...

Page 17: ...power cord IV Viewing tube assembly The field of view of the two eyes is different The interpupillar distance is not correct Adjust the interpupillar distance The dioptric correction is not right Adju...

Page 18: ...ther waste The separate collection of this equipment at the end of its lifetime is organized and managed by the producer The user will have to contact the manufacturer and follow the rules that he ado...

Page 19: ......

Page 20: ...s com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA Hungary hungary optikamicroscopes com DXPD...

Reviews: