Page 21
F
Fig. 32
ig. 32
F
Fig. 33
ig. 33
F
Fig. 34
ig. 34
10.13 Use of polarized light
1.
Move the polarizer “PO” into the light path until the slider
clicks into its position. (Fig. 32)
①
2.
Move the rotatable analyzer “AN” into the light path until the
slider clicks into its position. (Fig. 33)
3.
Put a flat mirror on the stage and get into approximate focus
the mirror surface.
4.
While observing into the eyepieces, rotate the scale of the
analyzer
①
until you achieve the darkest image. (Fig. 34)
•
Now the light extinction (so-called “crossed Nicol position”)
is achieved.
5.
Put a specimen on the stage, focus the specimen and begin
observation.