3.5
Camera settings and Microscopy
To have a good image of your sample or probe, camera settings need to be adjusted by the user in the
“Camera”
option in the main window (see
). One can select parameters to be adjusted
automatically by ticking ‘Automatic’ or
use a one-time automatic adjustment from the main software
window
“Auto Setup Camera”. The maximum ‘Frame rate’ of the camera is 47 fps and it is dependent on
the exposure time e.g. exposure time of 100 ms (0.1 s) limits the frame rate to 10 fps (1/0.1=10). To
achieve higher frame rates, set shorter exposure time. From the main window, one can select to turn on
3 different LED rings: Brightfield, Phase one and Phase two.
“LED intensity”
sliding bar in the main window
allows to adjust the intensity of LED rings.
For saving the images,
“
File format
”
of the “Camera”
window (see
) allows setting image saving
options
when using “
S
ave image“ button from the main window
(‘Front panel’)
or automatic saving of
images or videos
“D
uring experiments
”
. Images taken during the experiment will be saved in the same
location as the text files of indentation. Moreover, the objective type (4x, 10x, 20x and 40x) can be
selected in the Objective menu. If another objective is used, um/pixel value can be calibrated using XY
stage movement in the camera view.