background image

[Emissivity Ratio (DARK/BRIGHT mode)]

q

Enter the setting mode for "Emissivity Ratio (DARK/BRIGHT

mode)" according to the procedure described in page 11.

The initial setting is DARK (

ε

=0.95).

w

Press  

  or  

  button in the setting mode to change the

emissivity ratio (DARK or BRIGHT).

e

Press  SET/ENTER  button to set the emissivity ratio.

[Setting "˚C/˚F mode"]

q

Enter the setting mode for "˚C/˚F mode" according to the proce-

dure described in page 11.  The initial setting is ˚C.

w

Press  

  or  

  button in the setting mode to choose the tem-

perature display unit whether ˚C or ˚F mode.

e

Press  SET/ENTER  button to set the temperature display unit.

* When it is set for ˚F mode, the displayed temperature value

measured, set or stored in memory is automatically converted

to Fahrenheit in any mode.

[Record Erasing]

Note this function is to erase all the stored records in memory.

It is not able to erase data one by one.

q

Enter the setting mode for "Record Erasing" according to the

procedure described in page 11.

w

Press  

  or  

  button in the setting mode to select "CLR" sign on

the display.

e

Press  SET/ENTER  button, and all the recorded data are erased.

* If you wish to cancel the Record Erasing, press  

  button to select "ESC" sign on the

display and press  SET/ENTER  button.  Then, you can return to the setting mode again.

* Please note that OPTEX is not liable for any loss of data.

Emissivity (

ε

)

Emissivity is a value that indicates the infrared energy emitted from the surface of an object.  Every object has its own emissivity
value and it varies depending on the surface condition or the temperature of the object.  The emissivity ratio of PT-5LD is fixed at two
points, i.e., DARK (

ε

=0.95) and BRIGHT (

ε

=0.70).

Example : DARK (

ε

=0.95):  Food, rubber, plastic, painted area, etc.

BRIGHT (

ε

=0.70):  Oxidized metal, etc.

If the object has different emissivity value from either of the above ratio, there could be some possibilities that the measured
temperature value on the display shows different from the actual temperature value of the object.  Refer to the above examples as
a guideline for setting DARK/BRIGHT mode.  When you wish to measure shiny object like metals, put a piece of optional black tape
(

ε

=0.95) on the surface of the measuring object to cover the measuring area, if possible.

Setting 

––  "Emissivity Ratio (DARK/BRIGHT mode)", "˚C / ˚F mode" and "Record Erasing"

14

[BRIGHT mode]

"brt" and "0.70" are displayed in turn.

[DARK mode]

"drk" and "0.95" are displayed in turn.

[˚C mode]

[˚F mode]

[Record Erasing mode]

[Record Erasing Cancel mode]

Ramco National

800-280-6933 | [email protected] www.optex-ramco.com

All manuals and user guides at all-guides.com

Summary of Contents for PT-5LD

Page 1: ...81 77 579 8199 Take Care of the Environmet This manual uses recycled paper Printed in JAPAN 1023 3 2008 2 Portable Non Contact Thermometer Instruction Manual Ramco National 800 280 6933 nsales ramcoi...

Page 2: ...ce of an object into temperature This thermometer measures the surface temperature of solid and liquid without contacting them The temperature of gas cannot be measured by this thermom eter Contents I...

Page 3: ...s other than those specified herein may result in hazardous radiation exposure Safe Usage This instruction manual contains various warnings for your safety and proper usage to avoid possible personal...

Page 4: ...thermometer s sensitiv ity to emissivity is fixed at 0 95 0 70 the displayed temperature could differ from the actual temperature of objects which has different emissivity value See Emissivity on pag...

Page 5: ...opile Spectral response 8 14 m Response time 0 7 sec 90 Accuracy 0 95 0 200 C 2 C 201 C 1 Repeatability 1 C of reading value Display resolution 1 C Sighting method Non coaxial laser marker Class 2 HOL...

Page 6: ...on wakes up the thermometer The power supply turns off automatically after 15 seconds absence of pressing any button Name of Components 6 Ramco National 800 280 6933 nsales ramcoi com www optex ramco...

Page 7: ...7 Ramco National 800 280 6933 nsales ramcoi com www optex ramco com All manuals and user guides at all guides com...

Page 8: ...AX MIN HOLD See Setting on page 15 Continuous Measurement Starting Continuous Measurement qPress and hold both of MEASURE button and SET ENTER button simultaneously for two seconds or more then the Co...

Page 9: ...cle However please note that the measuring field of view enlarges in proportion to the measuring distance This thermometer has an optical resolution of 15 1 D Distance to the measuring object S Spot s...

Page 10: ...or more in HOLD Mode refer to page 8 and CALL sign appears on the LCD display In CALL recalling mode the last record of tem perature data and the corresponding memory number are shown on the display w...

Page 11: ...each setting is completed it automatically turns to the next setting mode For example when setting of High Limit Temperature for Alarm is completed it comes to the setting mode for Low Limit Temperat...

Page 12: ...perature than Low LimitTempera ture for Alarm Turning ON OFF Alarm Function qWhen the High Limit Temperature setting is completed the setting mode for ON OFF Alarm Function appears Press or button to...

Page 13: ...Attention The Low Limit Temperature for Alarm cannot be set at higher temperature than High LimitTempera ture for Alarm Turning ON OFF Alarm Function qWhen the Low Limit Temperature setting is complet...

Page 14: ...ign on the display and press SET ENTER button Then you can return to the setting mode again Please note that OPTEX is not liable for any loss of data Emissivity Emissivity is a value that indicates th...

Page 15: ...e field of view Adjust the measuring range referring to Field ofView shown in page 9 Affected by a nearby heating source Block the heating source with a shielding plate or something like that to avoid...

Page 16: ...spection Re calibration is recommended once a year Ask the distributor where you purchased the unit for details Battery Installing Battery When the battery level indicator shows little electricity lef...

Reviews: