4
2.4
Construction
Generator consist of modules connected together by DIN and/or AWP connectors.
•
main board
•
supply board
•
suppressor board
•
analog and digital signals controlling board
•
PWM board
•
filters board
•
power stage board
•
measurement board
•
RS 232 communication control board
2.5
Information plate
www.optel.pl
Power Ultrasound Generator
UG S/A
Serial no.: 07.12.rev.4
Date of production.: 03.2008
Ultrasound Frequency: 17 – 50kHz
Power Supply: 220-240 VAC/50Hz
Power Output: 600W
Dimensions: 360 x 340 x 160mm
Weight: app. 11 kg
Standard: 67/308/Ce
Fig. 2. Information plate
2.6
Technical information
•
frequency range:
17 000 - 50 000Hz
•
frequency setting resolution:
1Hz or 2Hz
•
maximum output power:
max 600W
•
maximum output voltage:
500V
sk
•
maximum output current:
4.0 A
sk
•
output voltage shape:
SIN (h max. < 10% )
•
accuracy of frequency measurement:
±
1Hz
•
accuracy of voltage measurement:
±
5%,
±
5 LSB
•
accuracy of current measurement:
±
5%,
±
5 LSB
•
accuracy of phase measurement:
±
2
°
(in range +50/-70
°
)
•
accuracy of absolute value of
o impedance measurement:
±
5% (in range 20-400
Ω
)
•
four temperature sensors:
PT100
•
load impedance (minimum):
40-300
Ω
1
•
supply
230 VAC / 50Hz
•
maximum output power:
600W
•
PC communication protocol:
RS232
•
three parts of load impedance range
•
adjustable output voltage in range:
1%-99% (with 1% resolution)
•
transformer output