CO

2

  / humidity and hand-held temperature measuring device CO

2

-TEMP(-RH)-HMG

2

 | 

4

  |  Sensors PG1  |  Data sheet No. 14108  | Version 11-2018  

LCD display

Symbols

TWA 

Time weighted average (8 hours)

STEL 

 Short-term exposure limit  
(15 minutes weighted average)

HOLD 

Readings are freezed unchanged 

MIN/MAX  

 Minimum/Maximun readings 

 

Low battery indicator

DP 

Dew point temperature (version -RH)

AIR 

Air temperature

WBT 

Wet bulb temperature (version -RH)

Unit of relative humidity

o

E (C/F) 

Celsius/Fahrenheit

Keypad

   Turns on and off the meter. Enters setup mode. Sets as 

non-sleep mode with 

.

   Exits setup page / mode. Enters CO

2

 calibration with 

   Freezes the current readings. Cancels data hold function.
   Activates or cancels the backlight. Selects unit or 

increases value in setup.

   Selects AIR, DP, WBT temps display. (version -RH only)

Selects unit or decreases value in setup.

   Activates MIN, MAX, STEL, TWA function. Saves and  

finishes settings.

 

Operation

Power ON/OFF

Press 

 to turn the meter on and off. At power up, it emits 

a short beep and performs 30 seconds countdown (Fig. 2) for 
meter warm up, then enters normal mode with current CO

2

, tem-

peratures, and humidity (version -RH) readings displayed (Fig. 3).

Taking measurement

The meter starts measurement when power on and update 
readings every second. In the condition of operating environ-
ment change (ex. from high to low temp.), it takes 30 sec to 
respond for CO

2

 sensor and 30 minutes for RH.

Note: 

Do not hold the meter close to faces because exhalation affects 
CO

2

 levels.

AIR (all models), DP, WBT (version -RH)

Press 

 to switch temperatures display. The lower left display 

will cycle from air temperature, dew point temp. (Fig. 4), and 
wet bulb temp. (Fig.5).

Data hold

Press 

 to freeze the readings, “HOLD” icon is displayed 

on the left top of LCD (Fig.6). All current readings are kept 
unchanged, except STEL and TWA. Press 

 again to cancel 

the hold function. 

Backlight

Hold down 

 for more than 1 second to activate and cancel 

backlight function.

MIN, MAX, STEL, TWA

Under normal mode, press 

 to see the minimum, maximum, 

and weighted average readings. Each press of 

, it displays 

MIN, MAX, STEL, TWA in sequence and returns to normal mode.

In MIN and MAX modes, it shows the minimum and maximum 
readings of CO

2

 on main display and of AIR or Dp or WB tempera-

tures and humidity (version -RH) on the lower displays. (Fig. 7)

In STEL and TWA modes, the main display shows the weighted 
average of CO

2

 readings for the past 15 minutes (STEL) and 8 

hours (TWA). The lower displays are the current AIR, DP/WB
temperatures and humidity (version -RH). (Fig. 8)

version -RH