16. EMC Declaration
Impulse Auto Mattress Instructions and Technical Specification
Declaration of Conformity
For EN 60601-1-2 (4th Ed.)
Company Name:
Opera Care
Company Address: Azure House, Connaught Road, HULL, HU7 3AP, UK
Trade Name:
OPERA
Model No.:
SR365 (Opera Flo), SR366 (Opera Flo Auto),
SR366 (Opera ECLIPSE Auto), SR366 (Opera IMPULSE Auto),
SR366 (Opera Flo Digital)
Report Number:
ETC 19-12-RBO-038
Power Supply:
AC 220 - 240V , 50/60Hz , 0.2A
Recommended separation distances between
portable and mobile RF communications equipment and the ME equipment
The Air Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Air Pump can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Air Pump as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=
1
5
,
3
V
P
80 MHz to 800 MHz
d=
1
5
,
3
E
P
800 MHz to 2.5 GHz
d=
1
7
E
P
2
.
0
1
.
0
1
.
0
1
0
.
0
7
.
0
4
.
0
4
.
0
1
.
0
3
.
2
2
.
1
2
.
1
1
4
.
7
7
.
3
7
.
3
0
1
3
.
3
2
7
.
1
1
7
.
1
1
0
0
1
Declaration – electromagnetic emissions
The Air Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Air Pump
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Air Pump uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Air Pump is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies