3

3.  Warnings

3.1  Siderails

Use this mattress with side rails that meet bed rail regulations (BS EN 60601-2-

52:2010) to ensure that the gap between the side rail and the top of the mattress 

is not large enough to pose risk of head or neck entrapment. Failure to do so could 

result in serious patient injury or death. If applicable, adhere to facility or local 

guidelines regarding entrapment regulations.

3.2  DO NOT Disassemble

contact Opera Care on 0333 222 8584 .

2.2  Use of Linens

It is recommended to limit bed linens to a single layer in order to allow moisture 

recommended for use with this mattress system. 

2.3  Flammability Hazard

exists. This device draws air from the surrounding environment. Thus, cigarette 

smoking may damage internal components.

2.4  Disinfection Between Patients

This system should be disinfected thoroughly between patients in order to avoid 

cross contamination.

2.5  Weight Capacity

Verify that the patient’s weight, accessories, and this mattress replacement system, 

do not exceed the bed frame’s manufacturer’s recommended weight capacity.

2.  Safety Precautions

2.1  Proper Operation

To ensure proper operation, please inspect and verify that all parts are set up 

properly and are anchored securely to the bed system. Verify that the mattress 

replacement system does not interfere with the bed frame function. Do not place 

anything on top of the power unit. Make sure power cord set is underneath bed

frame and does not pose a hazard.

Summary of Contents for Relieve Overlay

Page 1: ...Instructions and for Opera Relieve Overlay Mattress System Opera Care Azure House Connaught Road Hull HU7 3AP 0333 222 8584 support opera care co uk opera care co uk...

Page 2: ...ervice 11 14 Warranty 11 15 12 Thank you for purchasing this anti decubitus mattress replacement system PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE SETTING UP AND USING THE DEVICE Pay special atte...

Page 3: ...e ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area Always consult a physician or healthprofessional before using this mattress system This anti decubitus...

Page 4: ...Hazard exists This device draws air from the surrounding environment Thus cigarette smoking may damage internal components 2 4 Disinfection Between Patients This system should be disinfected thorough...

Page 5: ...l make sure to follow national state and or facility requirements for disposal of the unit 3 6 Maximum Operating Temperature The air mattress operates at a max of 100 F 37 7 C 3 7 Sharp Protrusions CA...

Page 6: ...e internal pressure of the mattress according to patient weight Simply turn the knob to adjust the pressure setting if the mattress is too soft or firm to suit each patient s needs Caregivers should a...

Page 7: ...ower Cord provided with a molded on anti kink bushing held in place by integral slot in bottom enclosure 5 Hanging Hooks Hanging hooks are designed to hang the power unit on almost any foot board 5 7...

Page 8: ...EN 60601 2 52 2010 6 2 Secure Power Unit Secure the power unit onto the foot board using hanging hooks 6 4 Connect Hose Couplers Firmly connect the air hose couplers to the couplers on the power unit...

Page 9: ...TUCKING THE BEDSHEET SHEET IN TOO TIGHTLY MAY REDUCE THE EFFECTIVENESS OF THE SYSTEM 8 Patient Transfer and Transport 8 1 Transfer It is recommended to have the mattress system fully inflated during...

Page 10: ...ying with a clean dry cloth Use only mild detergents to clean the coverlet and the mattress Any appropriate NON PHENOLIC cleaning agent may be used for heavy and coverlet are completely dry before let...

Page 11: ...ment is not and or Bottoming out is occurring and or Patient leaves a deep indentation at the contact area which does not return back to its original shape Check CPR Pull Cord Unzip mattress cover and...

Page 12: ...e production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by our company or if in the sole opinion of our company that the system sh...

Page 13: ...insulated IPX0 do not immerse power unit in any liquid or spray any liquids directly on the power unit This system is not AP APG Protected Continuous operation Power Cord Operating Transportation Sto...

Page 14: ...in which radiated RF disturbances are controlled The customer or the user of the Air Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF...

Page 15: ...1000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupt...