12. Troubleshooting

Fault

Inspection Procedure

Remedy

Power unit is not 

working.

plugged into both the control 

unit and the electrical outlet.

Secure power cord into control unit 

and/or electrical outlet.

Check if the power switch is 

in the ON position.

Turn power switch to ON position.

Check if power surge has 

shut down the power unit.

A power surge may overload the 

circuitry temporarily. Turn the unit 

off, and check the fuse for damage. 

Turn the unit on again with normal 

procedure.

Make sure there is no power 

failure.

Turn on and operate the unit after 

power is restored to the facility or 

home.

Power unit does not respond 

to possible solutions.

Please contact Opera Care on 0333 
222 8584 for assistance.

Power unit 

is working, 

but mattress 

replacement is not 

and/or

Bottoming out is 

occurring

and/or

Patient leaves a deep 

indentation at the 

contact area which 

does not return back 

to its original shape.

Check CPR Pull Cord

Unzip mattress cover and inspect 

CPR latch is fully inserted into 

socket.

Check if mattress’ air hose 

couplers are properly 

connected to power unit’s 

quick release couplers.

place.

Verify that patient weight 

setting is correct.

Increase or decrease weight 

setting until appropriate pressure is 

reached.

dirt.
Lift mattress coverlet up 

to check if air cells are 

connected correctly.

Make sure all air cells are properly 

linked to air supply.

Lift mattress coverlet up to 

check if air tubes are kinked 

or obstructed.

Check and adjust air tubes’ 

positions.

Check if air cells are cut or 

cracked.

Please contact Opera Care on 0333

Patient’s wounds 

are not responding 

to pressure relief 

(reddening of skin).

Contact your physician and/or 

nursing service immediately.

222 8584 for assistance.

Summary of Contents for Relieve Overlay

Page 1: ...Instructions and for Opera Relieve Overlay Mattress System Opera Care Azure House Connaught Road Hull HU7 3AP 0333 222 8584 support opera care co uk opera care co uk...

Page 2: ...ervice 11 14 Warranty 11 15 12 Thank you for purchasing this anti decubitus mattress replacement system PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE SETTING UP AND USING THE DEVICE Pay special atte...

Page 3: ...e ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area Always consult a physician or healthprofessional before using this mattress system This anti decubitus...

Page 4: ...Hazard exists This device draws air from the surrounding environment Thus cigarette smoking may damage internal components 2 4 Disinfection Between Patients This system should be disinfected thorough...

Page 5: ...l make sure to follow national state and or facility requirements for disposal of the unit 3 6 Maximum Operating Temperature The air mattress operates at a max of 100 F 37 7 C 3 7 Sharp Protrusions CA...

Page 6: ...e internal pressure of the mattress according to patient weight Simply turn the knob to adjust the pressure setting if the mattress is too soft or firm to suit each patient s needs Caregivers should a...

Page 7: ...ower Cord provided with a molded on anti kink bushing held in place by integral slot in bottom enclosure 5 Hanging Hooks Hanging hooks are designed to hang the power unit on almost any foot board 5 7...

Page 8: ...EN 60601 2 52 2010 6 2 Secure Power Unit Secure the power unit onto the foot board using hanging hooks 6 4 Connect Hose Couplers Firmly connect the air hose couplers to the couplers on the power unit...

Page 9: ...TUCKING THE BEDSHEET SHEET IN TOO TIGHTLY MAY REDUCE THE EFFECTIVENESS OF THE SYSTEM 8 Patient Transfer and Transport 8 1 Transfer It is recommended to have the mattress system fully inflated during...

Page 10: ...ying with a clean dry cloth Use only mild detergents to clean the coverlet and the mattress Any appropriate NON PHENOLIC cleaning agent may be used for heavy and coverlet are completely dry before let...

Page 11: ...ment is not and or Bottoming out is occurring and or Patient leaves a deep indentation at the contact area which does not return back to its original shape Check CPR Pull Cord Unzip mattress cover and...

Page 12: ...e production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by our company or if in the sole opinion of our company that the system sh...

Page 13: ...insulated IPX0 do not immerse power unit in any liquid or spray any liquids directly on the power unit This system is not AP APG Protected Continuous operation Power Cord Operating Transportation Sto...

Page 14: ...in which radiated RF disturbances are controlled The customer or the user of the Air Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF...

Page 15: ...1000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupt...

Page 16: ...Opera Care Azure House Connaught Road Hull HU7 3AP 0333 222 8584 support opera care co uk opera care co uk U 2S2301 30831 G V1 1...

Reviews: