Opera ECLIPSE AUTO Instructions Manual Download Page 3

 

operabeds.com

  |  

0333 222 8584

  |  

2

Indications

This air alternating mattress system is designed for patients who endure pressure 
ulcer and potential patients who wish to reduce the likelihood of pressure ulcer. 
This device is intended to treat and prevent pressure ulcers by facilitating blood 
circulation and decreasing pressure of each tissues contact area. Always consult 
a physician or healthcare professional before using this mattress system. This 
anti-decubitus mattress system is mainly for very high risk patient groups.

Thank you for purchasing this anti-decubitus mattress replacement 
system. 

Please read these instructions carefully before setting up and 

using the device. 

Pay special attention to the warnings and other 

safety information. Use of genuine components is essential for optimal 
performance. If you do not fully understand all the instructions, safety 
precautions, and warnings, do not use this device. If you have questions, 
please contact Opera on 0333 222 8584. 

 

i

Contradictions

Certain patient conditions are not suitable for using this type of device such as 
fracture of unstable vertebrae and illness of unstable vertebrae. Always consult a 
physician or health professional before using this device.

The use of this system does not replace the regular repositioning, monitoring 
and, nursing of the patient. 

Summary of Contents for ECLIPSE AUTO

Page 1: ...ctions and Technical Specifications V I S I T O P E R A B E D S C O M C A L L 0 3 3 3 2 2 2 8 5 8 4 Enhancing Lives Delivering Comfort O P E R A I M P U L S E A U T O H Y B R I D M A T T R E S S S Y S...

Page 2: ...autions 4 3 Warnings 5 4 System Package 6 5 Power Unit Features 7 6 Mattress and Power Unit Installations 9 7 Program Settings 10 8 Patient Transfer and Transport 11 9 Emergency CPR Deflation 11 10 Cl...

Page 3: ...ecubitus mattress replacement system Please read these instructions carefully before setting up and using the device Pay special attention to the warnings and other safety information Use of genuine c...

Page 4: ...he European Medical Device Directives 93 42 EEC amended by 2007 47 EC IP21 Protected against ingress of solid foreign objects 12 5mm diameter Protected against vertically falling water drops Double In...

Page 5: ...e layer in order to allow moisture to escape efficiently through the coverlet Breathable incontinent pads are recommended for use with this mattress system 2 3 Flammability Hazard Avoid using this dev...

Page 6: ...Not Disassemble Do not disassemble the power unit if you are not a qualified technician Please contact Opera on 0333 222 8584 3 3 AP APG Protection This product is not AP APG protected 3 4 Periodic R...

Page 7: ...Waste Do not fold or bend Not more than two layers should be stacked on top of each other For long term storage the power unit should be in a box or covered with a dust protector 3 10 Disposal of Was...

Page 8: ...rged The memory re call function recalls the previous setting when the pump is turned off on or after a power outage a b c d b Pressure Monitoring LED Power unit features an integrated pressure sensor...

Page 9: ...t is not bottoming out f Couplers Quick release female couplers are used to secure mattress air hoses to the power unit g Power Receptacle Insert power cord set firmly into receptacle h e f g e Rocker...

Page 10: ...Connect Hose Couplers Firmly connect the air hose couplers to the couplers on the power unit Remove existing mattress from the bed frame 6 2 Secure Mattress Replacement 6 6 Power Up Unit Plug the pow...

Page 11: ...e according to the patient sensation by using the comfort button on the control panel of the power unit If the patient feels the mattress is too soft or firm increase or decrease the mattress internal...

Page 12: ...unit s power cord from the wall outlet and the mattress should stay inflated for approximately 40 50 minutes This mattress system also has a safety foam base underneath the air cells to support the p...

Page 13: ...is not properly fitted to the mattress the hybrid alternating pressure cells may move There is also a high risk of cross contamination if the mattress is used without a coverlet Do not use electric o...

Page 14: ...pera on 0333 222 8584 for assistance Power unit is working but mattress replacement is not inflating and or Bottoming out is occurring and or Patient leaves a deep indentation at the contact area whic...

Page 15: ...ase inspect all accessories when you purchase our product If there is any damage or missing accessories when you receive the product please ask for a replacement from Opera on 0333 222 8584 within thr...

Page 16: ...ng of equipment or parts If the damage is a result from improper operation a reasonable service fee and part cost will be charged The warranty stated above is the only warranty made and is in lieu of...

Page 17: ...put 220 240Vac 50Hz 0 2A Power Consumption Normal Operation Max 20W Power Control Unit Weight 2 5 kg Fuse Rating T1A 250V Electrical Classification Class II Type BF applied part mattress IP Code IP21...

Page 18: ...17 Opera Beds Opera Impulse Hybrid Mattress System Instructions and Technical Specifications Version 2 Notes...

Page 19: ...operabeds com 0333 222 8584 18 Notes...