12. Troubleshooting

Fault

Inspection Procedure

Remedy

Power unit is not working.

Check if power cord is 

firmly plugged into both 

the control unit and the 

electrical outlet.

Secure power cord into 

control unit and/or electrical 

outlet.

Check if the power switch 

is in the on position.

Turn power switch to on 

position.

Check if the power surge 

has shut down the power 

unit.

A power surge may overload 

the circuitry temporarily. Turn 

the unit off, and check the 

fuse for damage. Turn the 

unit on again with normal 

procedure. 

Make sure there is no 

power failure.

Turn on and operate the unit 

after power is restored to the 

facility or home.

Power unit does not 

respond to possible 

solutions.

Please contact Opera on 

0333 222 8584 for assistance.

Power unit is working, but 

mattress replacement is 

not inflating

and/or

Bottoming out is occurring

and/or

Patient leaves a deep 

indentation at the contact 

area which does not return 

back to its original shape.

Check CPR Pull Cord.

Unzip mattress cover and 

inspect CPR latch is fully 

inserted into socket.

Check if mattress’ 

air hose couplers are 

properly connected to 

power unit’s quick release 

couplers.

Secure air hose couplers 

firmly into place.

Verify that patient weight 

setting is correct.

Increase or decrease weight 

setting until appropriate 

pressure is reached.

Inspect air filter for dust 

and dirt.

Clean or replace air filter.

Lift mattress coverlet up 

to check if air cells are 

connected correctly.

Make sure all air cells are 

properly linked to air supply.

Lift mattress coverlet up 

to check if air tubes are 

kinked or obstructed.

Check and adjust air tubes 

positions.

Check if air cells are cut or 

cracked

Please contact Opera on 

0333 222 8584 for assistance.

Summary of Contents for ECLIPSE AUTO

Page 1: ...ctions and Technical Specifications V I S I T O P E R A B E D S C O M C A L L 0 3 3 3 2 2 2 8 5 8 4 Enhancing Lives Delivering Comfort O P E R A I M P U L S E A U T O H Y B R I D M A T T R E S S S Y S...

Page 2: ...autions 4 3 Warnings 5 4 System Package 6 5 Power Unit Features 7 6 Mattress and Power Unit Installations 9 7 Program Settings 10 8 Patient Transfer and Transport 11 9 Emergency CPR Deflation 11 10 Cl...

Page 3: ...ecubitus mattress replacement system Please read these instructions carefully before setting up and using the device Pay special attention to the warnings and other safety information Use of genuine c...

Page 4: ...he European Medical Device Directives 93 42 EEC amended by 2007 47 EC IP21 Protected against ingress of solid foreign objects 12 5mm diameter Protected against vertically falling water drops Double In...

Page 5: ...e layer in order to allow moisture to escape efficiently through the coverlet Breathable incontinent pads are recommended for use with this mattress system 2 3 Flammability Hazard Avoid using this dev...

Page 6: ...Not Disassemble Do not disassemble the power unit if you are not a qualified technician Please contact Opera on 0333 222 8584 3 3 AP APG Protection This product is not AP APG protected 3 4 Periodic R...

Page 7: ...Waste Do not fold or bend Not more than two layers should be stacked on top of each other For long term storage the power unit should be in a box or covered with a dust protector 3 10 Disposal of Was...

Page 8: ...rged The memory re call function recalls the previous setting when the pump is turned off on or after a power outage a b c d b Pressure Monitoring LED Power unit features an integrated pressure sensor...

Page 9: ...t is not bottoming out f Couplers Quick release female couplers are used to secure mattress air hoses to the power unit g Power Receptacle Insert power cord set firmly into receptacle h e f g e Rocker...

Page 10: ...Connect Hose Couplers Firmly connect the air hose couplers to the couplers on the power unit Remove existing mattress from the bed frame 6 2 Secure Mattress Replacement 6 6 Power Up Unit Plug the pow...

Page 11: ...e according to the patient sensation by using the comfort button on the control panel of the power unit If the patient feels the mattress is too soft or firm increase or decrease the mattress internal...

Page 12: ...unit s power cord from the wall outlet and the mattress should stay inflated for approximately 40 50 minutes This mattress system also has a safety foam base underneath the air cells to support the p...

Page 13: ...is not properly fitted to the mattress the hybrid alternating pressure cells may move There is also a high risk of cross contamination if the mattress is used without a coverlet Do not use electric o...

Page 14: ...pera on 0333 222 8584 for assistance Power unit is working but mattress replacement is not inflating and or Bottoming out is occurring and or Patient leaves a deep indentation at the contact area whic...

Page 15: ...ase inspect all accessories when you purchase our product If there is any damage or missing accessories when you receive the product please ask for a replacement from Opera on 0333 222 8584 within thr...

Page 16: ...ng of equipment or parts If the damage is a result from improper operation a reasonable service fee and part cost will be charged The warranty stated above is the only warranty made and is in lieu of...

Page 17: ...put 220 240Vac 50Hz 0 2A Power Consumption Normal Operation Max 20W Power Control Unit Weight 2 5 kg Fuse Rating T1A 250V Electrical Classification Class II Type BF applied part mattress IP Code IP21...

Page 18: ...17 Opera Beds Opera Impulse Hybrid Mattress System Instructions and Technical Specifications Version 2 Notes...

Page 19: ...operabeds com 0333 222 8584 18 Notes...

Page 20: ...Impulse Hybrid Mattress System Instructions and Technical Specifications Version 2 Azure House Connaught Road Kingswood Hull HU7 3AP 0333 222 8584 support operabeds com operabeds com Version 2 Publis...

Reviews: