background image

 

HOBO External Temp/RH Data Logger (UX100-023A) Manual 

1-800-LOGGERS 

www.onsetcomp.com

 

the logger has a 10-minute logging interval and logged a 
data point at 9:05. Then, the high limit was surpassed 
and burst logging began at 9:06. Burst logging then 
continued until 9:12 when the sensor reading fell back 
below the high limit. Now back in normal mode, the next 
logging interval will be 10 minutes from the last burst 
logging point, or 9:22 in this case. If burst logging had not 
occurred, the next data point would have been at 9:15. 

 

A New Interval event will appear on the plot (if you select 
events for plotting in the Plot Setup window) each time 
the logger enters or exits burst logging mode. 

Statistics 

Statistics is a logging mode in which the logger calculates 
maximum, minimum, average, and standard deviation statistics 
during logging, recording the results at each logging interval 
based on samples taken at a rate you specify. This will result in 
up to four additional series per sensor that record the following 
information at each logging interval: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 
values. 

For example, let’s say both the temperature and RH sensors 
have been enabled, the logging interval is set to 5 minutes and 
the sampling interval is set to 30 seconds (with maximum, 
minimum, average, and standard deviation all enabled). Once 
logging begins, the logger will measure and record the actual 
temperature and RH sensor values every 5 minutes. In addition, 
the logger will take a temperature and RH sample every 30 
seconds and temporarily store them in memory. The logger will 
then calculate the maximum, minimum, average, and standard 
deviation using the samples gathered over the previous 5-
minute period and log the resulting values. When reading out 
the logger, this would result in 10 data series (not including any 
derived series, such as dew point): two sensor series (with 
temperature and RH data logged every 5 minutes) plus eight 
maximum, minimum, average, and standard deviation series 
(four for temperature and four for RH with values calculated 
and logged every 5 minutes based on the 30-second sampling). 

To set up statistics: 

1.

 

Select Statistics for Logging Mode in the Launch Logger 
window. If Statistics has already been configured for this 
logger, click the Edit button in the Launch Logger 
window. 

2.

 

Click the Maximum, Minimum, Average, and Standard 
Deviation checkboxes for each of the statistics you want 
to calculate during logging. Note that Average is 
automatically enabled when selecting Standard 
Deviation. 

Important:

 Statistics apply to all enabled 

sensors; every selected statistic will be calculated for all 
sensors (except battery voltage). For example, if both the 
temperature and RH sensors have been selected in the 
Launch Logger window and you select Average, then the 
average will be calculated for both temperature and RH. 
In addition, the more statistics you record, the shorter 
the logger duration and the more memory is required. 

 

3.

 

Set the sampling interval, which must be less than and a 
factor of the logging interval. Choose either a preset 
sampling interval or select Custom and enter your own 
sampling interval. Keep in mind that the more frequent 
the sampling rate, the greater the impact on battery life. 

4.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging 
Mode in the Launch Logger window to make additional 
changes. 

Once logging begins, click the Alarm/Stats button on the logger 
to cycle through the current maximum, minimum, average, and 
standard deviation data on the LCD screen. You can plot the 
statistics series once you read out the logger. 

Reading Out the Logger 

There are two options for reading out the logger: connect it to 
the computer with a USB cable and read out it with HOBOware, 
or connect it to a HOBO U-Shuttle (U-DT-1, firmware version 
1.18m030 or higher) and then offload the data files from the  
U-Shuttle to HOBOware. Refer to the HOBOware Help for more 
details. 

Recording Internal Logger Events 

The logger records the following internal events to track logger 
operation and status. You can plot these events in HOBOware 
after reading out the logger and opening the data file. 

Internal Event Name 

Definition 

Host Connected 

The logger was connected to the computer. 

Started 

The Start/Stop button was pressed to begin 
logging. 

Stopped 

The logger received a command to stop 
recording data (from HOBOware or by 
pushing the Start/Stop button). 

Button Up/Button 
Down 

The Start/Stop button was pressed for 
1 second. 

Chan <#> Alarm 
Tripped 

A sensor alarm has tripped; # is the channel 
or sensor number as shown in the Launch 
Logger window in HOBOware. 

Chan <#> Alarm 
Cleared 

A sensor alarm has cleared; # is the channel 
or sensor number as shown in the Launch 
Logger window in HOBOware. This event 
also contains the value that was furthest 
out of range for the sensor before the alarm 
cleared. 

New Interval 

The logger has entered or exited burst 
logging mode. 

Safe Shutdown 

The battery level dropped below 2.5 V; the 
logger performs a safe shutdown. 

Summary of Contents for Hobo UX100-023A

Page 1: ...ical Resolution 0 01 Response Time 15 seconds to 90 in airflow of 1 m s 2 2 mph Drift 1 per year typical Cable Length 1 83 m 6 ft Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0 to 95 RH...

Page 2: ...oximate battery power remaining The logger has been configured to stop logging when memory fills The memory bar indicates the approximate space remaining in the logger to record data When first launch...

Page 3: ...p logging and choosing a logging mode 1 Connect the logger and open the Launch Logger window To connect the logger to a computer use the USB cable provided Click the Launch icon on the HOBOware toolba...

Page 4: ...g Out the Logger for details Setting up Alarms You can set an alarm to trip on the logger when a sensor reading rises above or falls below a specified value To set an alarm 1 Click the Alarms button f...

Page 5: ...th burst logging You also cannot select the Stop Logging option Never wrap when full if burst logging is configured 2 In the Burst Logging window select a sensor In the example at right the temperatur...

Page 6: ...Standard Deviation checkboxes for each of the statistics you want to calculate during logging Note that Average is automatically enabled when selecting Standard Deviation Important Statistics apply t...

Page 7: ...if burst or statistics logging modes are active and battery performance A new battery typically lasts 1 year with logging intervals greater than 1 minute Deployments in extremely cold or hot temperat...

Reviews: