Onset HOBO MX CO2 Logger Manual Download Page 7

 

HOBO

 

MX

 

CO

2

 

Data

 

Logger

 

(MX1102A)

 

Manual

 

1

800

LOGGERS

 

7

 

www.onsetcomp.com

 

to

 

get

 

the

 

average

 

and

 

create

 

the

 

offset

 

from

 

400

 

ppm.

 

(See

 

the

 

steps

 

later

 

in

 

this

 

section

 

for

 

more

 

details

 

on

 

manual

 

calibration.)

 

This

 

is

 

recommended

 

if

 

the

 

logger

 

is

 

deployed

 

in

 

a

 

building

 

that

 

is

 

always

 

occupied,

 

if

 

you

 

want

 

the

 

logger

 

to

 

be

 

calibrated

 

more

 

frequently

 

than

 

every

 

eight

 

days

 

(the

 

normal

 

auto

 

calibration

 

schedule),

 

or

 

if

 

you

 

want

 

to

 

calibrate

 

the

 

logger

 

immediately

 

after

 

logging

 

begins.

 

Note:

 

Once

 

a

 

manual

 

calibration

 

is

 

performed,

 

the

 

24

hour

 

auto

 

calibration

 

is

 

canceled

 

and

 

an

 

auto

 

calibration

 

will

 

be

 

performed

 

eight

 

days

 

from

 

the

 

time

 

the

 

manual

 

calibration

 

occurred.

 

 

Auto

 

calibration.

 

Use

 

this

 

option

 

if

 

you

 

want

 

the

 

logger

 

to

 

automatically

 

calibrate

 

within

 

the

 

first

 

24

 

hours

 

after

 

logging

 

begins

 

and

 

then

 

every

 

eight

 

days

 

thereafter.

 

The

 

logger

 

will

 

be

 

calibrated

 

based

 

on

 

the

 

average

 

of

 

the

 

three

 

CO

2

 

measurements

 

that

 

follow

 

the

 

lowest

 

CO

2

 

value

 

identified

 

during

 

the

 

24

hour

 

or

 

8

day

 

time

 

period

 

as

 

applicable.

 

Important:

 

Accurate

 

auto

calibration

 

requires

 

the

 

building

 

or

 

location

 

where

 

the

 

logger

 

is

 

deployed

 

to

 

be

 

empty

 

at

 

least

 

once

 

during

 

the

 

eight

day

 

period

 

(for

 

example,

 

an

 

empty

 

office

 

building

 

during

 

the

 

weekend

 

or

 

overnight

 

will

 

typically

 

have

 

background

 

CO

2

 

levels

 

of

 

400

 

to

 

450

 

ppm).

 

If

 

the

 

logger

 

is

 

deployed

 

in

 

an

 

area

 

where

 

the

 

CO

2

 

level

 

does

 

not

 

go

 

down

 

to

 

400

 

ppm

 

during

 

the

 

eight

day

 

time

 

period,

 

then

 

manual

 

calibration

 

should

 

be

 

performed

 

regularly

 

instead

 

or

 

inaccurate

 

CO

2

 

readings

 

will

 

be

 

reported.

 

If

 

you

 

plan

 

on

 

using

 

auto

 

calibration

 

but

 

the

 

A

 

mabuilding

 

will

 

be

 

occupied

 

during

 

the

 

first

 

day

 

after

 

logging

 

begins,

 

then

 

you

 

can

 

use

 

the

 

manual

 

calibration

 

option

 

as

 

well.

 

You

 

can

 

manually

 

calibrate

 

the

 

logger

 

immediately

 

after

 

logging

 

begins

 

and

 

use

 

auto

 

calibration

 

thereafter.

 

Note:

 

Every

 

time

 

the

 

logger

 

is

 

started,

 

auto

 

calibration

 

will

 

occur

 

after

 

24

 

hours

 

and

 

then

 

again

 

after

 

eight

 

days

 

unless

 

a

 

manual

 

calibration

 

is

 

performed

 

first.

 

 

Altitude

 

compensation.

 

The

 

CO

2

 

sensor

 

must

 

compensate

 

for

 

locations

 

above

 

or

 

below

 

305

 

meters

 

(1,000

 

feet)

 

to

 

provide

 

an

 

accurate

 

reading.

 

You

 

must

 

enter

 

the

 

altitude

 

above

 

or

 

below

 

sea

 

level

 

in

 

either

 

meters

 

or

 

feet

 

when

 

configuring

 

the

 

logger

 

if

 

it

 

will

 

be

 

deployed

 

at

 

a

 

location

 

above

 

or

 

below

 

sea

 

level.

 

In

 

normal

 

use,

 

the

 

CO

2

 

measurement

 

will

 

vary

 

by

 

approximately

 

0.135%

 

of

 

the

 

reading

 

for

 

each

 

mbar

 

change

 

in

 

barometric

 

pressure

 

(the

 

sensor

 

is

 

calibrated

 

at

 

1,013

 

mbar).

 

Use

 

altitude

 

compensation

 

when

 

deploying

 

the

 

logger

 

for

 

the

 

best

 

CO

2

 

accuracy

 

possible.

 

To

 

access

 

calibration

 

and

 

altitude

 

compensation

 

settings

 

in

 

the

 

app:

 

1.

 

Tap

 

the

 

Devices

 

icon

 

(in

 

HOBOmobile,

 

tap

 

Loggers

 

at

 

the

 

top

 

of

 

the

 

screen).

 

Tap

 

the

 

logger

 

in

 

the

 

app

 

to

 

connect

 

to

 

it.

 

2.

 

Tap

 

Configure

 

in

 

HOBOmobile

 

or

 

tap

 

 

in

 

HOBOconnect.

 

3.

 

Tap

 

the

 

CO

2

 

sensor.

 

4.

 

Select

 

auto

 

calibration,

 

manual

 

calibration,

 

or

 

both.

 

5.

 

Select

 

“Altitude

 

Compensation”

 

and

 

enter

 

the

 

altitude

 

above

 

or

 

below

 

sea

 

level

 

where

 

the

 

logger

 

will

 

be

 

deployed

 

in

 

either

 

meters

 

or

 

feet.

 

In

 

HOBOmobile:

 

You

 

can

 

tap

 

Use

 

Location

 

if

 

you

 

want

 

the

 

location

 

services

 

within

 

the

 

phone

 

or

 

tablet

 

to

 

calculate

 

altitude.

 

6.

 

Tap

 

Done

 

or

 

Save.

 

7.

 

Tap

 

Start

 

to

 

load

 

settings

 

to

 

the

 

logger.

 

To

 

access

 

calibration

 

and

 

altitude

 

compensation

 

settings

 

in

 

HOBOware:

 

1.

 

Connect

 

the

 

logger

 

to

 

the

 

computer

 

with

 

the

 

USB

 

cable.

 

2.

 

From

 

the

 

Device

 

menu

 

in

 

HOBOware,

 

select

 

Launch.

 

3.

 

Click

 

the

 

CO

2

 

Settings

 

button.

 

4.

 

Select

 

auto

 

calibration,

 

manual

 

calibration,

 

or

 

both.

 

5.

 

Select

 

“Use

 

Carbon

 

Dioxide

 

sensor

 

altitude

 

compensation”

 

and

 

enter

 

the

 

altitude

 

above

 

or

 

below

 

sea

 

level

 

where

 

the

 

logger

 

will

 

be

 

deployed

 

in

 

either

 

meters

 

or

 

feet.

 

6.

 

Click

 

OK.

 

7.

 

Click

 

Start

 

in

 

the

 

Launch

 

Logger

 

window

 

to

 

load

 

settings

 

to

 

the

 

logger.

 

Note:

 

If

 

both

 

auto

 

calibration

 

and

 

manual

 

calibration

 

are

 

selected,

 

the

 

logger

 

will

 

automatically

 

calibrate

 

within

 

24

 

hours

 

after

 

logging

 

begins

 

unless

 

a

 

manual

 

calibration

 

occurs

 

during

 

that

 

time

 

period.

 

In

 

addition,

 

when

 

both

 

calibration

 

settings

 

are

 

selected,

 

the

 

eight

day

 

calibration

 

cycle

 

will

 

be

 

reset

 

any

 

time

 

a

 

manual

 

calibration

 

is

 

performed.

 

To

 

manually

 

calibrate

 

the

 

CO

2

 

sensor:

 

Important:

 

If

 

you

 

do

 

not

 

follow

 

these

 

manual

 

calibration

 

instructions

 

as

 

described,

 

the

 

sensor

 

readings

 

will

 

be

 

incorrect

 

and

 

you

 

will

 

need

 

to

 

manually

 

calibrate

 

the

 

logger

 

again.

 

1.

 

Take

 

the

 

logger

 

outside

 

in

 

fresh

 

air

 

on

 

a

 

dry

 

day

 

where

 

the

 

carbon

 

dioxide

 

level

 

is

 

400

 

ppm.

 

You

 

can

 

also

 

use

 

an

 

indoor

 

location

 

for

 

manual

 

calibration

 

if

 

it

 

is

 

unoccupied

 

and

 

is

 

not

 

exposed

 

to

 

a

 

ventilation

 

system.

 

2.

 

Press

 

the

 

Calibrate

 

button

 

on

 

the

 

logger

 

for

 

5

 

seconds

 

until

 

it

 

beeps.

 

The

 

logger

 

will

 

then

 

calibrate

 

for

 

5

 

minutes.

 

The

 

CO

2

 

and

 

Calibrate

 

symbols

 

on

 

the

 

LCD

 

will

 

flash

 

while

 

the

 

calibration

 

is

 

underway.

 

A

 

time

‐ 

and

 

date

stamped

 

manual

 

calibration

 

event

 

is

 

logged

 

in

 

the

 

data

 

at

 

the

 

end

 

of

 

the

 

5

minute

 

calibration

 

sequence.

 

3.

 

Once

 

the

 

Calibration

 

process

 

is

 

complete,

 

return

 

the

 

logger

 

to

 

its

 

deployment

 

location.

 

Repeat

 

this

 

process

 

at

 

least

 

once

 

every

 

eight

 

days

 

for

 

best

 

accuracy.

 

Setting

 

up

 

Alarms

 

You

 

can

 

set

 

an

 

alarm

 

to

 

trip

 

on

 

the

 

logger

 

when

 

a

 

sensor

 

reading

 

rises

 

above

 

or

 

falls

 

below

 

a

 

specified

 

value.

 

This

 

can

 

alert

 

you

 

to

 

problems

 

so

 

you

 

can

 

take

 

corrective

 

action.

 

To

 

set

 

up

 

a

 

sensor

 

alarm:

 

1.

 

In

 

the

 

app:

 

Connect

 

to

 

the

 

logger

 

and

 

tap

 

Configure

 

in

 

HOBOmobile

 

or

 

tap

 

 

in

 

HOBOconnect..

 

In

 

HOBOware:

 

From

 

the

 

device

 

menu,

 

select

 

Launch.

 

Click

 

the

 

Alarms

 

button

 

in

 

the

 

Launch

 

Logger

 

window.

 

2.

 

Select

 

the

 

sensor

 

that

 

you

 

want

 

to

 

set

 

up

 

with

 

an

 

alarm

 

condition.

 

3.

 

Enable

 

the

 

High

 

Alarm

 

if

 

you

 

want

 

an

 

alarm

 

to

 

trip

 

when

 

the

 

sensor

 

reading

 

rises

 

above

 

the

 

high

 

alarm

 

value.

 

Drag

 

the

 

slider

 

to

 

the

 

reading

 

that

 

will

 

trip

 

the

 

alarm

 

or

 

type

 

a

 

specific

 

reading.

 

Summary of Contents for HOBO MX CO2 Logger

Page 1: ...ccuracy 2 from 20 to 80 typical to a maximum of 4 5 including hysteresis at 25 C 77 F below 20 and above 80 6 typical Resolution 0 01 Drift 1 per year typical CO2 Sensor Range 0 to 5 000 ppm Accuracy...

Page 2: ...y life With app use battery life can be reduced by remaining connected excessive readouts audible alarms and paging Visual audible alarms and other events can have a marginal impact on battery life Me...

Page 3: ...rs there are the stronger the Bluetooth signal The logger is currently communicating with HOBOware via USB cable The logger is currently being powered by USB cable The battery indicator shows the appr...

Page 4: ...cted to the computer with the USB cable the LCD screen refreshes every second regardless of logging interval When the logger has stopped logging the LCD screen will remain on with STOP displayed until...

Page 5: ...d in HOBOmobile or and then in HOBOconnect Select this to create a password for the logger that will be required if another mobile device attempts to connect to it To reset a password simultaneously p...

Page 6: ...ing Mode Choose a logging mode Fixed Logging The logger will record data for all enabled sensors at the selected logging interval Burst Logging In burst mode logging occurs at a different interval whe...

Page 7: ...to it 2 Tap Configure in HOBOmobile or tap in HOBOconnect 3 Tap the CO2 sensor 4 Select auto calibration manual calibration or both 5 Select Altitude Compensation and enter the altitude above or below...

Page 8: ...ered by USB cable or every 5 minutes if it is powered by batteries If you are configuring a CO2 sensor alarm for a battery powered logger it is recommended that the duration you select is divisible by...

Page 9: ...logging mode See Logger Events for details on plotting and viewing the event In addition if the logger is stopped with a button push while in burst logging mode then a New Interval event is automatica...

Page 10: ...the app 1 Tap the Devices icon in HOBOmobile tap Loggers at the top of the screen Tap the logger in the app to connect to it 2 Tap Readout in HOBOmobile or tap in HOBOconnect 3 In HOBOmobile Tap the D...

Page 11: ...r operating range Expected battery life varies based on the ambient temperature where the logger is deployed the logging or sampling interval frequency of offloading and connection to a mobile device...

Page 12: ...any interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void...

Reviews: