18
Time Zone: Select Time Zone according to where you are.
Time: Set the current time.
Use 24 Hour Time Format
:
ON/OFF
About device
You can view your device information and details, including build version,w device hardware details, serial
number, and OS version.
Adaptor
Input: 100-240Vac, 50-60Hz, 1.1A
Output: 13.0Vdc, 3.69A, 47.97W
EMC Guidance
Warning: Don’t use near active HF surgical equipment and the RF shielded room of an ME system for
magnetic resonance imaging, where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
Warning: Use of accessories, transducers, and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 12 in (30 cm) to any part of the equipment, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity