79
6
Caring for your system
Cleaning your meter, lancing device
and cap
The meter, lancing device and cap should be cleaned
whenever they are visibly dirty and before disinfection.
Clean your meter at least once per week. For cleaning
obtain regular strength liquid dish soap and a soft cloth.
Prepare a mild detergent solution by stirring 2.5 mL of
regular strength liquid dish soap into 250 mL of water.
•
Do Not
use alcohol or any other solvent.
•
Do Not
allow liquids, dirt, dust, blood or control
solution to enter the test strip port or the data port.
(See page 16.)
•
Do Not
spray cleaning solution
on the meter or immerse it in
any liquid.
1. Holding the meter with the test strip port pointed
down, use a soft cloth dampened with water and mild
detergent to wipe the outside of the meter and lancing
device
Be sure to squeeze out any
excess liquid before you wipe
the meter. Wipe the outside of
the cap.
Summary of Contents for Vario Flex
Page 1: ...Blood Glucose Monitoring System Instructions for Use Owner s Booklet ...
Page 2: ......
Page 3: ...1 Owner s Booklet Blood Glucose Monitoring System ...
Page 17: ...15 1 Getting to know your system This page left blank intentionally ...
Page 19: ...17 1 Getting to know your system ...
Page 20: ...18 1 Getting to know your system Test strip Getting to know your OneTouch Verio Test Strip ...
Page 23: ...21 1 Getting to know your system This page left blank intentionally ...
Page 40: ...38 3 Taking a test Getting to know your OneTouch Delica Lancing Device ...
Page 87: ...85 7 Battery This page left blank intentionally ...
Page 113: ...111 Notes ...
Page 114: ...112 Notes ...
Page 115: ......