103
Detailed information about your system
9
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch Verio Flex™
System compares well with a laboratory method.
# of Subjects
# of Tests
Slope
Intercept
(mmol/L)
100
600
1.01
-0.02
95% CI
Slope
95% CI
Intercept
(mmol/L)
Std. Error
(S
y.x
)
(mmol/L)
R
2
1.00 to 1.01
-0.12 to 0.08
0.64
0.99
Summary of Contents for Vario Flex
Page 1: ...Blood Glucose Monitoring System Instructions for Use Owner s Booklet ...
Page 2: ......
Page 3: ...1 Owner s Booklet Blood Glucose Monitoring System ...
Page 17: ...15 1 Getting to know your system This page left blank intentionally ...
Page 19: ...17 1 Getting to know your system ...
Page 20: ...18 1 Getting to know your system Test strip Getting to know your OneTouch Verio Test Strip ...
Page 23: ...21 1 Getting to know your system This page left blank intentionally ...
Page 40: ...38 3 Taking a test Getting to know your OneTouch Delica Lancing Device ...
Page 87: ...85 7 Battery This page left blank intentionally ...
Page 113: ...111 Notes ...
Page 114: ...112 Notes ...
Page 115: ......