13
Changing the Insulin Cartridge
1.
Disconnect infusion set from your body.
2.
Remove infusion set from your skin and discard.
3.
Unscrew the cartridge compartment cap using a
counter-clockwise motion. Leave tubing
connected to the cartridge.
4.
Pull cartridge straight out.
5.
Disconnect tubing from cartridge and discard
cartridge.
6.
Fill new cartridge according to Instructions for
Use included with your cartridge.
7.
Remove infusion set tubing cap from Luer
connector. (Not all infusion sets have these
caps.)
8.
Thread the Luer connector of the infusion set
through the smaller opening of the cartridge
compartment cap.
9.
Remove cap from the filled cartridge tip.
10.
Attach infusion set Luer connector to
cartridge tip using clockwise motion until
snug. Then twist another quarter turn.
Changing the Insulin Cartridge and Priming Your Pump
Correctly threaded through cap
Connect infusion set
to cartridge here
Luer connector
of infusion set
Top of cartridge
compartment cap
Close
Open
OneTouch
®
Ping
®
Insulin Pump
I